- 14 -
Secondly, select “
YES
” for “
START TESTING
” on OSD “DICOM CONFORMANCE
TEST”.
※
If “DICOM GSDF” is NOT selected, the message “START TESTING” will not
appear.
2 M P
D I C O M C O N F O R MA N C E
T E S T
R E S U L T
:
- - -
MA X D E V I A T I O N : - - -
L U M I N A N C E :
S t a b l e
E L A P S E D T I ME :
1 2 3 m i n
S T A R T T E S T I N G
Y e s
> N o
DICOM CONFORMANCE TEST
START TESTING
Completing the measurement takes between 1 minute to 2 minutes.
8)
Showing the test results
When the measurement is completed, the test
results are shown.
2 M P
D I C O M C O N F O R MA N C E
T E S T
R E S U L T
:
P a s s e d
MA X D E V I A T I O N : 4 . 5
L U M I N A N C E :
S t a b l e
E L A P S E D T I ME :
1 2 3 m i n
S T A R T T E S T I N G
Y e s
> N o
Test results
Test results
Passed: Max.
deviation ≤ 10%
Failed: Max. deviation > 10%
Cancelled: Test has been
cancelled.
Passed(!), Failed(!
)
When “Passed” or “Failed” is
followed by “(!)”, the test has
been made under not
recommended condition.
(The test was performed in
less than 60 minutes of
continuous lighting time of
the backlight.)
If the test is continuously
performed in less than 30
seconds, “Unstable” is shown
in LUMINANCE and “(!)” is
shown in RESULT.
Maximum deviation
The test results are saved until the next test or
calibration.
If the achieved maximum deviation does not
conform to the standard even though the test
has been performed under the recommended
condition, calibrate the monitor using an
external calibration sensor.
Without the OSD operation for over 15 seconds,
the OSD disappears automatically.