Operating instructions K-Laser Blue Dental Rev. ZE
Page 16 of 72
4.3 Power supply
Connect the power cable to the DC IN socket at the back of the device.
Caution
The device may only be operated with the Sinpro MPU101-106 power supply. Operation with
other power supplies may result in failure or destruction of the laser unit. If any power supply
other than the one recommended is used, the approval of the entire unit automatically
becomes void and the warranty expires.
Warning
The use of any power supplies other than the one recommended may cause overheating and
failure of the laser unit or damage of batteries.
The device is supplied with a rechargeable battery and therefore can be used without connected
power cable. The status of the rechargeable battery and whether the power cable is actually
connected will be always displayed on the touch screen.
Notice
There will be a warning if the rechargeable battery will reach a low level of capacity.
Notice
While operating, should the charge level of the battery be below 5%, the green led will flash at
5 “ intervals to remind the user to plug in the unit.
The device is fully functional and can be run while charging the battery.
Charge the battery completely.
Notice
The rechargeable battery is automatically charged, also during the device use, in the event of
power supply connection.
Notice
The rechargeable battery must be fully charged regularly. After six months the rechargeable
battery might lose its loading capacity and it may not be rechargeable anymore.
Notice
To have a precise indication of the battery charge status, please perform a periodic battery
calibration process every 3 months. See chapter “Set up” for further details.
4.3.1 Replacing the rechargeable battery
If the rechargeable battery does not charge more than 30% the battery needs to be replaced.
Contact the Authorized Service Center.
Caution
Only use the specific battery pack provided by Eltech K-Laser s.r.l. (see chapter spare parts
and optional).