Operating instructions K-Laser Blue Dental Rev. ZE
Page 19 of 72
Contact surgical handpiece body assembly
The operator can attach the handpiece body (B) to the connector (A). When attaching or
detaching the parts, we recommend to slightly rotate the two parts. This guarantees an easier
connection. The user must hear a sound (click) locking parts. This guarantees a correct
connection.
Attach the handpiece body (B) to the connector (A). The pin must be completely in the groove.
Disposable tip assembly
Attach the disposable tip (D) firmly on the sleeve (C) prior to mounting the optic fiber tip (E).
Caution
Check whether the disposable tip is damaged before attaching it. Replace it if necessary.
Caution
Please check that the disposable tip is firmly attached prior to each use.
Warning
Always use the disposable tip with the fiber and check the correct fixation.
Warning
To avoid the risk of damaging optic fiber tip use only provided disposable tips.
Optic fiber assembly
Remove the protective cap from the optic fiber tip and preserve it for the fiber sterilization.
Remove the protective cap from the handpiece body and preserve it.
Insert the optic fiber tip into the handpiece body and screw it firmly.
Warning
Slightly tighten the optic fiber tip so that it can be easily loosened. Make sure not to overwind!
Caution
Never use the laser without fiber, check for correct fixation.