Operating instructions K-Laser Blue Dental Rev. ZE
Page 31 of 72
The device will automatically define the following parameters:
The explanation of the impulse profile on the screen is described into more details.
The impulse is displayed in light blue. Horizontal line and
numerical value on the right represent the average power that the
device automatically calculates.
If ISB modality is on, a bar with red profile (970nm emission) is
present on the base among the impulses.
If ISB modality is off the bar doesn't appear.
It allows to select wavelengths, in single or combined modality.
445nm selected: the impulse is light blue.
970nm selected: the impulse is yellow.
445nm + 970nm selected: the impulse is violet.
The setting of the impulse can be performed in under described modalities.
Pulse modality
. Sets the number of the impulses. It can be defined by pressing
numeric keyboard or by
-/+
key to increase or decrease the number of impulses.
Frequency modality
. Sets the emission time expressed in seconds.
It can be defined by the user by pressing the numeric keyboard or by
-/+
key to
increase or decrease the number of seconds. The 0 value on numeric keyboard
corresponds to a laser emission in an infinite sequence. This is shown as
.
CW modality
. Where predicted. Sets the emission time expressed in seconds in
continuous wave (CW). In this modality Ton is automatically set in
CW
and Toff
in
--
. The 0 value on the numeric keyboard corresponds to a laser emission in an
infinite sequence. This is shown as
.
Number of the impulses to be shot or emission time.
Press the numeric keyboard or
-/+
key set the desired value.
The device calculates the energy delivered according to selected
power and time.
The device shows the energy emitted during treatment.
The duty cycle is defined as the ratio between the pulse duration
(when the laser beam is actually actuated) and the total period of
time (which is the time from the beginning of a pulse to the
beginning of the next pulse).
The device calculates the frequency of the impulse based on the
Ton
– Toff by the operator.
Frequency (Hz)
Energy
Applied
Duty Cycle (%)