Operating instructions K-Laser Blue Dental Rev. ZE
Page 33 of 72
After selection of the following area
Dermatologic surgery
the following screen is displayed.
Different levels of the red aiming beam can be
defined.
Graphic of the laser impulse.
Wavelength selector. Press it to select needed
wavelengths, in individual or combined way.
Treatment phases.
if available.
Press to increase or decrease the value.
ISB modality. It is on if the green bar is present.
DISB modality. It is on if the green bar is present.
Press it to save a treatment.
Press it to go back one screen.
Press it to enter a manual screen. User confirmation
is needed.
Laser armed. After a delay of 2 seconds by this
button pressure, the aiming beam is switched on and
the laser is ready to be used.
The device will automatically define the following parameters:
The device calculates the energy delivered according to selected
power and time.
The device shows the energy emitted during treatment.
The duty cycle is defined as the ratio between the pulse duration
(when the laser beam is actually actuated) and the total period of
time (which is the time from the beginning of a pulse to the
beginning of the next pulse).
The device calculates the frequency of the impulse based on the
Ton
– Toff by the operator.
Frequency (Hz)
Energy
Applied energy
Duty Cycle (%)