Operating instructions K-Laser Blue Dental Rev. ZE
Page 41 of 72
To manage the parameters select the field that will be highlighted with a green square and then
press:
Minus / Plus key or
to enter in a Keyboard to implement the modification.
The following graphic helps the user to understand the shape of the impulse to be set.
Ton is the time during which the laser is emitting.
Toff is the time of the impulse during which the laser is off.
The height of the impulse is the peak power of emission.
The yellow line and the numerical value represent the
average power that the device automatically calculates on
the base of Ton, Toff and power set by the operator. This
line gets on and off based on the average power.
Warning
In order to increase the energy and thus the tissue cutting strength, it is recommended to
decrease the
T-OFF
time, rather than to increase the
T-ON
time. In this way the energy of each
impulse remains the same but
the impulses’ frequency increases. So the thermal damage
around the treated tissue is limited.
Notice
Press the key
to save treatment. The saved treatment is stored in the Manual area.
Notice
You can rename an already stored treatment. Press the key
.
Notice
Treatments can be removed from the list. In the Manual area press the
.
Setting can start from a minimum of 0,1 W to the max power
delivered by the device. It can be set by step of 0.01W pressing
-/+
or by numeric keyboard.
Time (in ms) during which the laser is emitting.
It is adjustable between CW, 0.1
– 999 ms.
Time (in ms) during which the laser is off.
It is adjustable between 0.1
– 999 ms.
Power
T - ON
T - OFF