Operating instructions K-Laser Blue Dental Rev. ZE
Page 47 of 72
Warning
All persons present in the room must wear the laser protective goggles as soon as it is
advised during use.
Notice
Before starting a laser treatment in battery operation please reconfirm the battery status.
The following fields allow the operator to manage the treatment parameters.
Selecting this field and pressing the numeric keyboard is possible to modify the
parameters. It is also possible to set parameters by touching - /+.
Selecting this field and pressing the numeric keyboard is possible to add a new
phase (the new one is created and it is positioned after the selected one) or
delete the current one. The treatment is composed up to twelve phases.
The selected wavelength is highlighted in a green field.
The not selected wavelength is highlighted in a grey-green field.
On the event the operator needs to modify and adjust the power, the
selected wavelength is highlighted in a red field.
Each wavelength can be singularly selected.
For each treatment phase, the device keeps trace of the performed modifications. The followings
images show the colors used for the status identification.
In this screen:
The operator is working in this phase No modifications have been
performed respect to the already present values.
The phase is present but not currently used. No modifications have
been performed respect to the already present values.
The phase is present but not currently used. At least one parameter
has been modified respect to the already present values.
The following steps explain the parameters management.
Select
and select the phase to be modified.
Select the needed parameters.
To manage the parameters select the field that will be highlighted with a green square and then
press:
Minus / Plus key or
to enter in a Keyboard to implement the modification.
Peak
Frequency
Phase time
Phase