Operating instructions K-Laser Blue Dental Rev. ZE
Page 18 of 72
Warning
Never direct the laser beam toward a person’s eye or thyroid gland. All persons present in the
room (patient, operator and assistant) must always wear the laser protective goggles delivered
along with the device.
Warning
Before and during the emission laser verify that the aiming beam is present and that it projects
a regular shape. If it does not come projected some shape, or if it is much irregular one, the
optic fiber or the device could be damaged. In this case, proceed as follows:
Switch off the laser and check the optic fiber as well as the fiber connection for mechanical
damage;
If the optic fiber is damaged;
If you cannot detect any damage on the optic fiber and the signal of the aiming laser is not
visible, switch off the laser and contact Authorized Service Center.
4.4.1 Contact surgical handpiece
This handpiece is made of the following parts:
A) Optic fiber with connector
B) Contact surgical handpiece body
C) Sleeve for contact surgical handpiece body
D) Disposable tip
E) Optic fiber tip
Caution
Prevent dust, dirt and foreign particles from entering the handpiece.
Make sure that it is always cleaned: before, during and after the use.
Caution
Always cover the connectors with the special protective caps provided for this purpose after the
removal of the handpiece.
Caution
The disposable tip is supplied non-sterile. It must be sterilized before use.
Initial check
Warning
Every time before using the handpiece, make sure that the optic fiber tip, the disposable tip, the
bending tool and the sleeve have been disinfected and sterilized in autoclave.
Notice
The optic fiber tip, the disposable tip, the bending tool and the sleeve can be sterilized in an
autoclave with saturated water vapor at minimum sterilization values of 134°C (273.2°F), 3 min.
holding time and 2.04 bar (29,59 psi) overpressure without damage.