Operating instructions K-Laser Blue Dental Rev. ZE
Page 52 of 72
5.7.3 Set touch
This key allows the operator to set the switch modality.
Single
touch
Laser active if fingerswitch is kept pressed.
Double
touch
Laser is active at the first press of the fingerswitch.
Second press terminates the emission.
Remote
Restart
Therapy
Remote Restart Control. Perform a treatment again
using the button on the handpiece.
Press it to select the finger switch (optional).
Press it to select the wireless footswitch.
Operator goes back one screen.
Press it and follow the instruction on screen to set the
pairing to the footswitch.
Press it to confirm the switch modality.
Notice
The green bar shows the selected modality
Notice
Single touch.
If the switch is kept pressed, the message “LASER ON” appears and at the
same time, yellow LEDs light up, audible alarm sounds and laser starts emitting. At the switch
release, the laser is deactivated, but remains ready for operation.
Notice
Double touch
.
At the first pressure of the switch, the message “LASER ON” appears, at the
same time, yellow LEDs light up, audible alarm sounds and laser starts emitting. At the switch
release, the laser is still active. Press the switch a second time to deactivate the laser.
Notice
In Surgical area the device automatically disables Double touch modality.
Notice
The
RRT
(
R
emote
R
estart
T
herapy) modality allows to perform again a complete treatment
without removing the hands from the handpiece. It is sufficient to press the fingerswitch one
time to close the “completed therapy” window.
In
Double touch
, the second pressure of the fingerswitch enables the termination of the
laser beam.
In
Single touch
modality excludes
RRT
and vice versa.