Operating instructions K-Laser Blue Dental Rev. ZE
Page 58 of 72
7.2 Cleaning of the contact surgical handpiece
Warning
Before cleaning, switch off the device and disconnect the power cable from the power supply.
Optic fiber with connector
Contact surgical handpiece body
Sleeve for contact surgical handpiece body
Optic fiber tip
Caution
Prevent dust, dirt and foreign particles from entering the handpiece.
Make sure that the optic system is always cleaned: before, during and after the use.
Caution
Always cover the connectors with the special protective caps provided for this purpose after the
removal of the handpiece.
You can disinfect it using any of the products that are commonly used to disinfect medical electrical
equipment, e.g. MinutenWipes from Alpro and Caviwipes TM.
Observe the instructions provided by the manufacturers of these disinfectants.
Notice
All debris must be removed from the optical fiber tip before it is detached from the handpiece.
Notice
Use only not spray disinfectants that comply with the requirements of your national authorities
and whose bactericidal, fungicidal and virucidal properties have been tested and properly
certified.
Notice
Do not use spray disinfectant or insert disinfectant on the handpiece optics!
Notice
Cover with the proper protective caps the connectors.
Notice
Clean the optic with a dry, soft cloth to remove dust from the optics. More stubborn spots can
be removed with a damp cloth. Please proceed carefully not to scratch and damage the parts.