Operating instructions K-Laser Blue Dental Rev. ZE
Page 62 of 72
7.5 Cleaning of no contact surgical handpiece
Warning
Before cleaning, switch off the device and disconnect the power cable from the power supply.
Optic fiber with connector
No contact surgical handpiece
body
Set of no contact surgical tip
Caution
Prevent dust, dirt and foreign particles from entering the handpiece.
Make sure that the optic system is always cleaned: before, during and after the use.
Caution
Always cover the connectors with the special protective caps provided for this purpose after the
removal of the handpiece.
You can disinfect it using any of the products that are commonly used to disinfect medical electrical
equipment, e.g. MinutenWipes from Alpro and Caviwipes TM.
Observe the instructions provided by the manufacturers of these disinfectants.
Notice
Do not use spray disinfectant on the handpiece optics!
Notice
Use a gauze soaked with no alcohol detergent to clean the handpiece optics. Clean and
disinfect the no contact surgical tips.
Notice
Use only disinfectants that comply with the requirements of your national authorities and whose
bactericidal, fungicidal and virucidal properties have been tested and properly certified.
Notice
Clean the optic with a dry, soft cloth to remove dust from the handpiece.
Please proceed carefully not to scratch and damage the parts.
Sterilization
Sterilize the set of no contact surgical tip.
Notice
The set of no contact surgical tip can be sterilized in an autoclave with saturated water vapor at
minimum sterilization values of 134°C (273.2°F), 3 min. holding time and 2.04 bar (29,59 psi)
overpressure without damage.