There is no aiming beam or it does not project a uniform circular pattern.
Check if the optic fiber is damaged. If the optic fiber is damaged, replace it with a new one.
Check if the connection of the optic fiber to the handpiece is damaged.
Wireless foot control is not working.
Check the battery of the wireless footswitch.
Check the registration of the wireless footswitch.
Check if the wireless footswitch is chosen in the set-up submenu.
If you cannot solve the malfunctioning, switch off the laser and contact an Authorized Service
Center.
8.3 Safety check
The following safety checks must be performed every 24 months by a qualified service engineer.
Visual inspection of the unit and its accessories for mechanical damage that might impair
operation;
General function check;
Check of the visual and audible indicators;
Check the electrical safety;
Check the power emission of the laser;
Check the cleaning of the optics.
8.4 Technical support, repair and testing
In case technical support, contact the following address:
ELTECH K-LASER s.r.l.
Via Castagnole, 20/H,
31100 Treviso, Italy
Phone +39 0422 210 430
Fax +39 0422 297 137
Email service@klaser.it
Before sending the device to the Service Center, it is required to disinfect the device and
accessories as written in the present manual.
The device may be sent in for repair or for safety inspection only in its original packaging, including
all accessories.
Eltech K-Laser s.r.l. is the only provider of technical service for its products, unless otherwise
indicated by the company. This is to guarantee the product’s safety and for the preservation of the
product’s characteristics.