Operating instructions K-Laser Blue Dental Rev. ZE
Page 65 of 72
9
Manufacturer’s declaration on electromagnetic
compatibility
9.1 Definitions
Notice
The device complies with all requirements for electromagnetic compatibility according to IEC
60601-1-2.
9.1.1 Emission (electromagnetic)
When electromagnetic energy is emitted by a source.
9.1.2 Interference immunity
The ability of a device or system to work without errors even if there is electromagnetic
interference.
9.1.3 Immunity level
The maximum level of a certain electromagnetic interference that affects a particular device or
system, where the device or system remains operative with a certain level of performance.
ELECTROMAGNETIC EMISSION
EMISSION TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT -
GUIDANCE
RF emissions
CISPR 11
Group 1
The unit uses RF energy only for its internal
function. The RF emission is therefore very
low and it is improbable that nearby
electronic devices might be disturbed.
RF Emission
CISPR 11
Class A
The unit is suitable for use in all
establishments, except for those destined to
domestic use and those directly connected to
the public, low-voltage power supply network,
that supplies establishments destined for
residential use.
Warning:
The equipment is
intended for use by healthcare professionals
only. This equipment may cause radio
interference or may disrupt the operation of
nearby equipment. It may be necessary to
take mitigation measures, such as re-
orienting or relocating the equipment or
shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies