Operating instructions K-Laser Blue Derma Rev. ZE
Page 20 of 70
4.4.2 Contact surgical handpiece
This handpiece is made of the following parts:
A) Optic fiber with connector
B) Contact surgical handpiece body
C) Sleeve for contact surgical handpiece body
D) Disposable tip
E) Optic fiber tip
Caution
Prevent dust, dirt and foreign particles from entering the handpiece.
Make sure that it is always cleaned: before, during and after the use.
Caution
Always cover the connectors with the special protective caps provided for this purpose after the
removal of the handpiece.
Caution
The disposable tip is supplied non-sterile. It must be sterilized before use.
Initial check
Warning
Every time before using the handpiece, make sure that the optic fiber tip, the disposable tip, the
bending tool and the sleeve have been disinfected and sterilized in autoclave.
Notice
The optic fiber tip, the disposable tip, the bending tool and the sleeve can be sterilized in an
autoclave with saturated water vapor at minimum sterilization values of 134°C (273.2°F), 3 min.
holding time and 2.04 bar (29,59 psi) overpressure without damage.
Contact surgical handpiece body assembly
The operator can attach the handpiece body (B) to the connector (A). When attaching or
detaching the parts, we recommend to slightly rotate the two parts. This guarantees an easier
connection. The user must hear a sound (click) locking parts. This guarantees a correct
connection.
Attach the handpiece body (B) to the connector (A). The pin must be completely in the groove.
Disposable tip assembly
Attach the disposable tip (D) firmly on the sleeve (C) prior to mounting the optic fiber tip (E).
Caution
Check whether the disposable tip is damaged before attaching it. Replace it if necessary.
Caution
Please check that the disposable tip is firmly attached prior to each use.