Operating instructions K-Laser Blue Derma Rev. ZE
Page 21 of 70
Warning
Always use the disposable tip with the fiber and check the correct fixation.
Warning
To avoid the risk of damaging optic fiber tip use only provided disposable tips.
Optic fiber tip assembly
Remove the protective cap from the optic fiber tip and preserve it for the fiber sterilization.
Remove the protective cap from the handpiece body and preserve it.
Insert the optic fiber tip into the handpiece body and screw it firmly.
Warning
Slightly tighten the optic fiber tip so that it can be easily loosened. Make sure not to overwind!
Caution
Never use the laser without fiber, check for correct fixation.
Insert the handpiece body into the sleeve making sure that the optic fiber tip easily slides.
Slide the sleeve to regulate the optic fiber tip. The optic fiber tip must protrude at least 1 cm out
of the disposable tip.
Warning
If the optic fiber tip does not protrude at least 1 cm out of the disposable tip, there is a risk that
the disposable tip will heat up.
Warning
In the event of the sleeve is left inserted for a long time, it could happen that it is "glued" to the
gasket. If it happens, move the sleeve with more strength till the complete removal of the part.