Operating instructions K-Laser Blue Derma Rev. ZE
Page 27 of 70
4.6 Interlock
As foreseen by the normative the device is supplied with connector interlock for the automatic
interruption of the emission laser at distance.
Such connector is normally connected to a switch which interrupts the functioning of the laser in
case a person enters in the room in which it is used.
Notice
The installation must be performed by a qualified electrician who is also responsible for the
installation and maintenance of the electrical system to which the device is connected.
Notice
Please request the technical data sheet with wiring diagram for the installation of the interlock
device.
Notice
Additional or different safety precautions required by the applicable national or local regulations
for the protection of operators, assistant personnel, or patients must also be observed.
4.7 Wireless footswitch
The wireless footswitch must be assigned to the device via a registration.
To register the footswitch, proceed as follows:
In “Set-up” select the item “Set-touch”.
Press the key to start the procedure.
Follow the on-screen instructions.
Caution
Touch a grounded metal part before opening the housing to prevent damage to the PC board
due to electrostatic discharge.
Caution
Prior to changing the batteries, switch the device off at the main switch. This prevents
accidental triggering.
Removing and replacing the batteries:
Remove the screws from the bottom of the footswitch.
Remove the cover and replace the battery (2 AA type cells).
Be careful to insert it with the correct polarity (minus pole facing spring).
Place the battery holder back again in the compartment and close it with the cover.
Screw the screws at the bottom of the footswitch.
Notice
After changing the batteries, switch the device and check the complete functionality of the
footswitch. It is not necessary to register again the footswitch again at the device after
changing batteries.