Operating instructions K-Laser Blue Derma Rev. ZE
Page 33 of 70
After selection of the following areas:
Vascular surgery
,
Dark spots
and
Podiatry
the following
screen is displayed.
Different levels of the red aiming beam can be
defined.
Graphic of the laser impulse.
Wavelength selector. Press it to select needed
wavelengths, in individual or combined way.
Energy of each emitted impulse.
Speed: number of impulses emitted in a second.
ISB modality. It is on if the green bar is present.
Press it to save a treatment.
Press it to go back one screen.
Press it to enter a manual screen. User confirmation
is needed.
Laser armed. After a delay of 2 seconds by this
button pressure, the aiming beam is switched on and
the laser is ready to be used.
Where predicted the device allows to directly enter into a manual area.
The device will automatically define the following parameters.
The explanation of the impulse profile on the screen is described into more details.
The impulse is displayed in light blue. Horizontal line and
numerical value on the right represent the average power that the
device automatically calculates.
If ISB modality is on, a bar with red profile (970nm emission) is
present on the base among the impulses.
If ISB modality is off the bar doesn't appear.
It allows to select wavelengths, in single or combined modality.
445nm selected: the impulse is light blue.
970nm selected: the impulse is yellow.
445nm + 970nm selected: the impulse is violet.
The device shows the energy emitted during treatment.
The duty cycle is defined as the ratio between the pulse duration
(when the laser beam is actually actuated) and the total period of
time (which is the time from the beginning of a pulse to the
beginning of the next pulse).
Applied energy
Duty Cycle (%)