Operating instructions K-Laser Blue Derma Rev. ZE
Page 40 of 70
When the desired values are defined, the user can:
press the button to save treatment.
Device asks to insert the name.
Confirm the action.
The treatment is saved into Manual area.
or
press
to start working. The laser is armed:
Device warns to wear protective goggles.
After confirmation the green LEDs start lighting.
After a delay of 2 seconds the aiming beam is switched on.
The laser is now ready for operation.
Notice
The user can disable the sound pressing the sound icon present on the laser activation screen.
The sound is disabled for the current treatment.
Warning
All persons present in the room must wear the laser protective goggles as soon as it is advised
during use.
Notice
Before starting a laser treatment in battery operation please reconfirm the battery status.
When you actuate the foot switch the laser starts emitting. At the same time, two yellow LEDs at
the upper right and left end of the device control unit light up as well as the laser active bar on the
touch screen and the audible alarm sounds. When you release the foot switch to interrupt
treatment, the laser is deactivated, but remains ready for operation.