Operating instructions K-Laser Blue Derma Rev. ZE
Page 46 of 70
The operator
can modify
cannot modify
The device automatically manages the following parameters:
Treatment total time. This value is automatically calculated by
the device.
The device automatically calculates the energy delivered
during treatment (in J) based on the power settings and the
time.
The device automatically calculates the average power based
on power and duty cycle.
Notice
Treatment time and Total Joule are automatically calculated by the device and they cannot be
set by the user. The other parameters can be modified.
The operator can select the following areas to manage a treatment:
It is the peak power and it is adjustable between 0.1 W and
max value of the model. In 0.01 W increments by numeric
keyboard or by touching
.
Notice
For the use of optics, please read carefully specific
instructions and warnings.
The device calculates the average power from the power
values and duty cycle.
Frequency can be set in an infinite modality by selecting CW
or may be adjusted from 1 - 10000 Hz by numeric keyboard
or touching
1 Hz at a time.
The time of each phase is adjustable between 1 to 999 s by
numeric keyboard or by touching
The treatment is composed up to twelve phases.
Select the phase icon and touch the keyboard to add a new
phase or delete the current one.
Wavelengths are selectable to be used one at a time or
combined.
Frequency
Phase
(Peak) Power
Wavelength
Average power
Phase time
Treatment time
Total Joule
(Peak) Power
Frequency
Wavelength
Phase time
Average power
Phase
Average (power)