Operating instructions K-Laser Blue Derma Rev. ZE
Page 47 of 70
For the treatment management:
Set the parameters for the first phase and proceed for the maximum number of 12 phases.
To delete completely a phase, select it and then press delete phase.
To confirm the treatment press the save button and insert the name.
After pressing the confirmation button, the device will store the treatment.
Press
to arm the laser.
Device warns to wear protective goggles.
After confirmation the green LEDs start lighting.
After a delay of 2 seconds, the aiming beam is switched on.
The laser is now ready for operation.
Notice
Before the laser emission starts, the device will ask the operator to verify the optics in use.
When you actuate the switch the laser starts emitting. At the same time, two yellow LEDs at the
upper right and left end of the device control unit light up as well as the laser active bar on the
touch screen and the audible alarm sounds. When you release the switch to interrupt treatment,
the laser is deactivated, but remains ready for operation.
Notice
The user can disable the sound pressing the sound icon present on the laser activation screen.
The sound is disabled for the current treatment.
Notice
In base of the selected frequency (CW or chopped modality), the aiming beam emits in a
different way, as described:
CW
: the aiming beam is fixed, not flashing;
Modulated
: the aiming beam is slowly flashing;
The following explanation describes treatment saving options:
Before treatment, you can press the button
and enter in a menu that
allows you to:
- save this treatment.
Moreover it is always possible to modify the parameters in order to customized
the treatment. Press the button
to store modifications.