Operating instructions K-Laser Blue Derma Rev. ZE
Page 60 of 70
7.5 Cleaning of the ENT handpiece
Warning
Before cleaning, switch off the device and disconnect the power cable from the power supply.
Caution
Prevent dust, dirt and foreign particles from entering the handpiece.
Make sure that the optic system is always cleaned: before, during and after the use.
Caution
Always cover the connectors with the special protective caps provided for this purpose after the
removal of the handpiece.
You can disinfect the optic using any of the products that are commonly used to disinfect medical
electrical equipment, e.g. MinutenWipes from Alpro and Caviwipes TM.
Observe the instructions provided by the manufacturers of these disinfectants.
Notice
Use only disinfectants that comply with the requirements of your national authorities and whose
bactericidal, fungicidal and virucidal properties have been tested and properly certified.
Notice
Clean the optic of the handpiece with a dry, soft cloth. More stubborn spots can be removed
with a damp cloth. Please proceed carefully not to scratch and damage the optic.
Sterilization
Sterilize the ENT handpiece.
Notice
The ENT handpiece can be sterilized in an autoclave with saturated water vapor at minimum
sterilization values of 134°C (273.2°F), 3 min. holding time and 2.04 bar (29,59 psi)
overpressure without damage.