K205ULT Operating Manual

Setting the Temperature

• Press Set key for 3 seconds. it display the code St. 
• Press Set key again to display the temperature SETPOINT. 

Modify by pressing the UP or DOWN keys. 

Changing Parameters - Enter Admin Section

• Press Set key for 3 seconds, it will display the code St. 
• Press DOWN arrow to Po. Press Set - 00.
• Press DOWN arrow to password (55). Press Set. 
• Choose parameter by scrolling with UP or DOWN arrows. 

Select parameter by pressing Set. 

• Change parameter settings with the UP and DOWN arrow
• When finished changing parameters press Set.

Parameter

Description

Setting scope

Default

User menu

St

Set cabinet temperature point

LS to US

-80

Administrator menu

Po

Administrator menu password

(The password is 55 and 

cannot be changed)

55

C8

Cabinet temp. probe calibration when below 0°

Use this parameter to match temp display with datalogger

-15.0~15.0

0

C9

Condenser probe calibration (High-temp. compressor loop)

-15.0~15.0

0

C10

Ambient temp. probe calibration

-15.0~15.0

0

C11

High-temp. evaporator probe calibration

-15.0~15.0

0

C13

The Minimum set temp. of cabinet

-95.0~C14

-86

C14

The Maximum set temp. of cabinet

C13~50.0

-40

C15

Maximum temp. of high-temp. evaporator when low-temp. 

compressor turned off

C7~10.0

-12

H1

The cycle time of capillary heating wire

1~300

108

H2

The working time of capillary heating wire

1~90

3

H3

Maximum delay time of capillary heating

1~90

60

A3

Alarm output delay after opening cabinet door

0~60

1

A4

Alarm ringback time when the alarm is not cleared

0~60

10

A5

Maximum temp. of over-temp. alarm of ambient (A5>A6)

A6~70.0

40

A6

Minimum temp. of over-temp. alarm of ambient (A6<A5)

-10.0~A5

5

A7

Delay time of over-temp. alarm of ambient

0~60

10

do1

Door switch controls output

0:Cancel door switch 

1:Alarm when the door open

0

do2

Whether a buzzer response is required when the door is open

0:No   1:Yes

1

Cd1

High temperature alarm starting value of condenser

30.0~Cd3               

38

Summary of Contents for K205ULT

Page 1: ...ay damage the unit or void your warranty Please contact K2 before attempting to change advanced parameters Unfamiliar with the operation of a K2 controller Use the video tutorials on our website resources page or call us for assistancee with special parameters Some K2 units can be changed from Celsius to Fahrenhiet We do not recommend changing your unit to Far henheit Your controller has operation...

Page 2: ... must be performed by qualified service personnel Do not use this appliance for other than its intended use Do not cover the front grille or block the rear air entry by placing object up against the cabinet Ensure adequate ventilation If the main supply cord is damaged it must be replaced by the manufacturer its service agent or similarly qualified and skilled persons in order to avoid hazard Do n...

Page 3: ... to prevent any overflow Ventilation All models must have clear and unobstructed ventilation from the entire surface area of the front grille Temperature Controller 3 SET RST Enter a parameter setting Switch between menu and parameter Adjust menu and parameters After 3s parameter setting downloads Adjust menu and parameters After 3s parameter setting uploads Exit from parameter settings Press for ...

Page 4: ...ssor stop High temperature compresso time delay Low temperature compressor work in non forced refrigeration mode Low temperature compressor is closed in non forced refrigeration mode Forced refrigeration mode Capillary heating starts Capillary heating is closed High temperature alarm of secondary system condenser High temperature protection of secondary system condenser Door open alarm Grid power ...

Page 5: ...gger 15 0 15 0 0 C9 Condenser probe calibration High temp compressor loop 15 0 15 0 0 C10 Ambient temp probe calibration 15 0 15 0 0 C11 High temp evaporator probe calibration 15 0 15 0 0 C13 The Minimum set temp of cabinet 95 0 C14 86 C14 The Maximum set temp of cabinet C13 50 0 40 C15 Maximum temp of high temp evaporator when low temp compressor turned off C7 10 0 12 H1 The cycle time of capilla...

Page 6: ...K205ULT Operating Manual Product Description 5 ...

Page 7: ...K205ULT Operating Manual Maintenance 6 Alarm Code ...

Page 8: ...1 Contact us via our chat feature at www k2sci com 2 Email support k2sci com 3 Call 800 218 7613 When storing vaccines you may be required to preform a field validation test A NIST callibrated external data logger may be used for this purpose A logger with text email or online access is an added layer of protection for your product load in the event of a temperature excursion K2 offers NIST calibr...

Page 9: ...50 0 40 C15 Maximum temp of high temp evaporator when low temp compressor turned off C7 10 0 12 H1 The cycle time of capillary heating wire 1 300 108 H2 The working time of capillary heating wire 1 90 3 H3 Maximum delay time of capillary heating 1 90 60 A3 Alarm output delay after opening cabinet door 0 60 1 A4 Alarm ringback time when the alarm is not cleared 0 60 10 A5 Maximum temp of over temp ...

Reviews: