background image

K205ULT Operating Manual

Maintenance

6

Alarm Code

Summary of Contents for K205ULT

Page 1: ...ay damage the unit or void your warranty Please contact K2 before attempting to change advanced parameters Unfamiliar with the operation of a K2 controller Use the video tutorials on our website resources page or call us for assistancee with special parameters Some K2 units can be changed from Celsius to Fahrenhiet We do not recommend changing your unit to Far henheit Your controller has operation...

Page 2: ... must be performed by qualified service personnel Do not use this appliance for other than its intended use Do not cover the front grille or block the rear air entry by placing object up against the cabinet Ensure adequate ventilation If the main supply cord is damaged it must be replaced by the manufacturer its service agent or similarly qualified and skilled persons in order to avoid hazard Do n...

Page 3: ... to prevent any overflow Ventilation All models must have clear and unobstructed ventilation from the entire surface area of the front grille Temperature Controller 3 SET RST Enter a parameter setting Switch between menu and parameter Adjust menu and parameters After 3s parameter setting downloads Adjust menu and parameters After 3s parameter setting uploads Exit from parameter settings Press for ...

Page 4: ...ssor stop High temperature compresso time delay Low temperature compressor work in non forced refrigeration mode Low temperature compressor is closed in non forced refrigeration mode Forced refrigeration mode Capillary heating starts Capillary heating is closed High temperature alarm of secondary system condenser High temperature protection of secondary system condenser Door open alarm Grid power ...

Page 5: ...gger 15 0 15 0 0 C9 Condenser probe calibration High temp compressor loop 15 0 15 0 0 C10 Ambient temp probe calibration 15 0 15 0 0 C11 High temp evaporator probe calibration 15 0 15 0 0 C13 The Minimum set temp of cabinet 95 0 C14 86 C14 The Maximum set temp of cabinet C13 50 0 40 C15 Maximum temp of high temp evaporator when low temp compressor turned off C7 10 0 12 H1 The cycle time of capilla...

Page 6: ...K205ULT Operating Manual Product Description 5 ...

Page 7: ...K205ULT Operating Manual Maintenance 6 Alarm Code ...

Page 8: ...1 Contact us via our chat feature at www k2sci com 2 Email support k2sci com 3 Call 800 218 7613 When storing vaccines you may be required to preform a field validation test A NIST callibrated external data logger may be used for this purpose A logger with text email or online access is an added layer of protection for your product load in the event of a temperature excursion K2 offers NIST calibr...

Page 9: ...50 0 40 C15 Maximum temp of high temp evaporator when low temp compressor turned off C7 10 0 12 H1 The cycle time of capillary heating wire 1 300 108 H2 The working time of capillary heating wire 1 90 3 H3 Maximum delay time of capillary heating 1 90 60 A3 Alarm output delay after opening cabinet door 0 60 1 A4 Alarm ringback time when the alarm is not cleared 0 60 10 A5 Maximum temp of over temp ...

Reviews: