FOR SAFETY MEASUREMENTS!!
WARNING
INTRODUCTION
2. SPECIFICATIONS
3. SAFETY PRECAUTIONS
4. NAME ILLUSTRATION
WARNING 1. Checks of Body and Test Lead
WARNING 3.
Warning for High Voltage Measurement
s
WARNING 4. Dangerous Voltage Measurement Procedures
WARNING 2. High Power Line Measurements Prohibited
WARNING 7. Test Lead Detachment
WARNING 6. Maximum Measurement Range Observance
WARNING 5. Correct Selection of RANGE Switch
1
2
3
4
To prevent an electrical shock hazard to the operator and/or damage to the
instruments, read this instruction manual carefully before using the
Multitester. WARNINGS with the symbol on the Multitester and this
instruction manual are highly important.
Important Symbols
The symbol listed in IEC 61010-1 and ISO 3864 means "Caution
(refer to instruction manual)".
WARNING :
The symbol in this manual advises the user of an
electrical shock hazard that could result in serious injury or even
death.
CAUTION :
The symbol in this manual advises the user of an
electrical shock hazard that could cause injury or material damages.
Do not measure High Power Line that might exceed 2kVA. Measurement on
High Power Line is very dangerous. It sometimes includes High Surge
Voltage that could cause dangerous arcs of explosive short in the instrument
and could result in serious injury to the operator. When measuring high
voltage, even for Low Power Line (Low Energy Circuit), take extra care to
avoid electrical shock hazard and/or damage to the instrument.
Thank you for purchasing KAISE "MODEL KF-20 ANALOG MULTITESTER". To
obtain the maximum performance of this instrument, read this Instruction Manual
carefully, and take safe measurement.
Inspect the instrument and acessories for transport damage. If there is any
damage or missing items, ask your local dealer for replacement.
Confirm that the following items are contained in the package.
2-1. MEASUREMENT SPECIFICATIONS
2-2. GENERAL SPECIFICATIONS
3-1. WARNINGS
3-2. GENERAL WARNINGS AND CAUTIONS
1. Analog Multitester
2. Test Lead (100-51)
3. Holster
4. Spare Battery (1.5V R6P)
5. Instruction Manual
Allligator
Clip
Red
Test Lead
Black
Test Lead
Allligator
Clip
Red
Test Lead
Black
Test Lead
Children and the persons who do not have enough knowledge
about electric measurements must not use this instrument.
Do not measure the electricity naked or barefooted to protect
yourself from electrical shock hazard.
Be careful not to get hurt with the sharp test lead pins.
Do not polish the case or attempt to clean it with any cleansing
fluid like gasoline or benzine. If necessary, use silicon oil or
antistatic fluid.
Avoid the multitester from hard mechanical shock or vibration,
high temperature and strong magnetic field.
Remove the batteries when the multitester is out of use for a
long time. The exhausted batteries might leak electrolyte and
corrode the inside.
WARNING 1.
WARNING 2.
WARNING 3.
CAUTION 1.
CAUTION 2.
CAUTION 3.
Measurement exceeding the specified maximum values of each measurement
range is prohibited.
Always confirm that
RANGE
Switch is set to the correct position. Do not measure
voltage in DC-mA,
Ω
and BAT. ranges.
Detach test leads from the measuring circuit before rotating
RANGE
Switch or
opening rear case for battery or fuse replacement.
WARNING
Detach test leads from the circuit before rotating
RANGE
Switch to avoid electric
shock hazard and serious damage to the instrument.
Select suitable scale from "0 - 6", "0 - 12", or "0 - 30" applying certain multiple.
Reading examples :
Multiply the scale value depending on the measurement range.
Reading examples :
Read the scale directly.
For accurate measurement, read following notes carefully.
DC 0.3V range :
Read "0 - 30" scale multiplying by 1/100.
DC 120V range :
Read "0 - 12" scale multiplying by 10.
DC 60uA range :
Read "0 - 6" scale multiplying by 10.
DC 300mA range :
Read "0 - 30" scale multiplying by 10.
"x 1" range :
Read the scale directly.
"x 100" range :
Multiply the scale value by 100.
"x 1k" range :
Multiply the scale value by 1,000.
Correct knowledge of electric measurements is essential to avoid unexpected
danger such as operator's injury or damage to the instrument. Read carefully and
observe the following precautions for safety measurements.
Before measurement, confirm the body of this instrument and handle insulators of
the Test Lead have no cracks or any other damages. Dust, grease and moisture
must be removed.
This instrument is applicable for Low Energy Circuit measurements only.
Do not use for High Power Line (High Energy Circuit) measurements such as
Distribution Transformers, Bus Bars and Large Motors.
Even for Low Energy Circuits of electric/electronic appliances, such as heating
elements, small motors, line cords and plugs, High Voltage Measurements are
very dangerous. Do not touch Multitester, Test Leads, and any part of the circuit.
Generally, shock hazard could occur when the current between the circuit, that
involves more than 33V rms or 46.7V DC or peak, and ground goes up to 0.5mA
or more.
For dangerous voltage measurement, strictly observe the warnings below.
(refer to fig.1)
ANALOG MULTITESTER
KF-20
INSTRUCTION MANUAL
In case of live-line measurement, strictly observe the warnings
below (refer to fig 2.)
fig. 1
fig. 5
fig. 2
fig. 3
fig. 4
1. UNPACKING AND INSPECTIONS
1 pce.
1 set
1 pce.
1 pce.
1 pce.
Turn off the circuit to be measured.
Plug black test lead to "- (COM)" terminal and red test lead into "+" terminal.
Attach black and red alligator clips to test lead pins.
Set
RANGE
Switch to desired DC or AC voltage range.
Connect black alligator clip to - (earth) side and red alligator clip to + (positive)
side of the measuring circuit.
Do not hold Multitester and test leads in your hands. Keep safety distance
from power source or circuit to be measured not to touch the dangerous
voltage.
Turn on the circuit to be measured, and read the meter.
Turn off the circuit again. Detach alligator clips after confirming the meter
returns into "0".
1.
2.
3.
4.
5.
6.
7.
8.
Set
RANGE
Switch to desired DC or AC voltage range.
Attach alligator clip to black test lead pin, and connect to - (earth) side of the
measuring circuit.
Do not hold Multitester in your hands. Keep safety distance from power source
or circuit to be measured not to touch the dangerous voltage.
Connect red test lead to + (positive) side, and read the meter.
Detach red test lead from the circuit first, then detach black alligator clip.
1.
2.
3.
4.
5.
Before measurement, confirm that meter points to "0". (zero-adjusted)
Select proper range referring to
「
5-1. 6-(1) Range Selection
」.
Read the meter from directly over the pointer.
Measurement in strong magnetic field or on iron plate causes reading error or
meter sensitivity wrong.
Once-a-year periodical calibration is necessary to maintain tolerance accuracy
and to take safety measurements.
1.
2.
3.
4.
5.
Set the switch to a desired measurement range.
To measure uncertain voltage or current, firstly measure the approximate
value at the highest range, then change the switch into the suitable range.
For correct voltage and current measurements, select the range to point the
right side of the meter scale. (between center scale and maximum scale)
For correct resistance measurement, select the range to point around the
center of
OHMS
scale.
a.
b.
c.
DC Voltage ( DC-V )
AC Voltage ( AC-V )
DC Current ( DC-mA )
Resistance (
Ω
)
1.5V Battery Test
Hi Range
Lo Range
Temperature (
℃
)
DC :
±
3
%
f.s., AC :
±
4
%
f.s.
Ω
/ TEMP :
±
3
%
f.s.length
1.5V R6P (AA) battery
×
1
136(H)
×
90(W)
×
30 (D)mm, 215g
100-51 Test Lead
×
1 set
Holster
×
1
1.5V R6P (AA) Battery (built-in and spare)
×
1 each
Spare Fuse (0.3A/250V)
×
1 (inside the rear case)
Instruction Manual
660 Clamp Adapter
810-02 Temperature Probe
940 Alligator Clip
0.3 / 3 / 12 / 30 / 120 / 300 / 1200V
12 / 30 / 120 / 300 / 1200V
60 A / 3mA / 30mA / 300mA
5K
Ω
/ 500K
Ω
/ 5M
Ω
1.5V R20P / R14P / R6P
1.5V 250mA
1.5V 50mA
−
50
℃〜+
250
℃
μ
Accuracy
Overload Protection
Power Supply
Dimensions & Weight
Accessories
Optional Accessories
×
1 and 300mA range : protected by internal diode
and 0.3A / 250V fuse
1200
1200
300
120
30
12
300
120
30
12
3
3
60
30 300
0.3
Hi
Lo
BAT.
×
1 ×100 ×1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC
DC
AC
BATT
TEMP
1.5V VOLT
BATT
m
A
MADE IN JAPAN
NO.
PROTECTION
10,000
/
V AC
20,000
/
V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
‑50
1.5
1.0
0.5
0
℃
CAL
OHM
S
KF-20
C O M
Ω
R E
P L
A C
G O O
D
Meter Case
Pointer
Zero Adjustment
Screw
RANGE Switch
Rear Case
Meter Scale
0
Ω
ADJ. Knob
+
Terminal
−
( COM)
Terminal
100-51 Test Lead
KF-20 contains 1 x fast-acting 0.3A/250V (
φ
5x20mm) fuse to protect Resistance
and Current measurement ranges. For replacement, refer to
「
6-2. FUSE
REPLACEMENT
」.
Insert black test lead into - (COM) terminal, and red test lead into + terminal.
Always confirm that
RANGE
Switch is set to the proper measurement range. To
avoid serious damage to the instrument, do not measure voltage in DC-mA,
Ω
and BAT ranges.
3. FUSE
Before measurement, confirm that meter points to "0". If not, adjust the pointer by
Zero Adjustment Screw
to take accurate measurement.
4. ZERO ADJUSTMENT
5. TEST LEAD
7. Meter Scale Reading
8. Measurement Accuracy
DC/AC scale
(for AC/DC Voltage, DC Current measurements : V,
μ
A, mA)
OHM scale
(for Resistance measurement :
Ω
)
TEMP scale
(for Temperature measurement :
℃
)
(1) Range Selection
(2) Confirmation of the Range
6. RANGE SWITCH
Read INSTRUCTION MANUAL carefully to understand the specification and
functions correctly.
「
3. SAFETY PRECAUTIONS
」
is higly important for safety
measurement.
KF-20 contains 1 x 1.5V R6P (AA) battery. Replace it referring to
「
6-1.
BATTERY REPLACEMENT
」
when exhausted.
5-1. PREPARATION FOR USE
1. INSTRUCTION MANUAL
2. BATTERY
5. MEASUREMENT PROCEDURES
TEMP
DC
5k
2k
1k
AC
DC
AC
BATT
TEMP
1.5V VOLT
BATT
NO.
PROTECTION
10,000
/
V AC
20,000
/
V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CA
L
OHM
S
R E
P L
A C E
G O O
D
UCHIDA
BATT scale
TEMP scale
DC/AC scale
OHMS scale
TAUT BAND
DROP
PROOF
1200
1200
300
120
30
12
300
120
30
12
3
3
60
30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC
DC
AC
BATT
TEMP
TAUT BAND
1.5V VOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000
/
V AC
20,000
/
V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
25
0
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CA
L
OHM
S
C O M
Ω
R E
P L
A C E
G O O
D
KF-20
1200
1200
300
120
30
12
300
120
30
12
3
3
60
30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC
DC
AC
BATT
TEMP
TAUT BAND
1.5V VOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000
/
V AC
20,000
/
V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CA
L
OHM
S
C O M
Ω
R E
P L
A C E
G O O
D
KF-20
outlet
when measuring DC. mA /
Ω
range
1200
1200
300
120
30
12
300
120
30
12
3
3
60
30 300
0.3
Hi
Lo
BAT.
×
1
×
100
×
1k
AC-V
DC-V
μ
DC
TEMP
TEMP
DC
5k
2k
1k
AC
DC
AC
BATT
TEMP
TAUT BAND
1.5V VOLT
BATT
m
A
MADE IN JAPAN
NO.
DROP
PROOF
PROTECTION
10,000
/
V AC
20,000
/
V DC
20
10
5
30
15
6
3
50
10
4
2
100
200
5
2
1
500
0
0
0
20
8
4
25
10
5
30
12
6
250
200
100
50
0
-50
1.5
1.0
0.5
0
℃
CA
L
OHM
S
C O M
Ω
R E
P L
A C E
G O O
D
KF-20
×
DROP
PROOF