EIASMANUAL 0918
KAM CONTROLS, INC.
15
M A I N T E N A N C E C O N T I N U E D
28.
29.
23.
24.
25.
26.
27.
22.
FIG. 4-11
On the remaining sampling section, remove the Bottom Seal Holder by using a 1 1/8" wrench. (Fig. 4-10)
FIG. 4-10
1 1/8" Wrench Flat
Bottom Seal Holder
Cover
Plunger Seal
Spring and
Check Valve
Bottom
Seal Holder
Using a 5/32" Allen wrench remove the Cover hold-
ing the plunger seal. (Fig. 4-11)
Inspect the Spring and Check Valve for damage and
replace if necessary.
Using solvent clean the inside threads of the Cover.
Place Check Valve and Spring back in place.
Install a new Plunger Seal.
Clean the outside Cover threads and apply a small
amount of blue Loctite 242 or equivalent.
Screw the cover back in place using 5/32" Allen
wrench
NOTE:
Do not alter small set screw on flat of Bot-
tom Seal Holder threads. This determines sample
bite volume. To change, please consult factory.