KAM CONTROLS, INC.
27
CALIBRATION EXAMPLE
Manual samples water content:
1. 2.51%
2. 2.75%
3. 2.62%
Average = 2.63%
LRW screenshots water content:
1. 2.60%
2. 2.91%
3. 2.70%
Average = 2.74%
Difference of averages:
2.74% – 2.63% = 0.11%
According to the averages, the LRW is reading 0.11% high.
Current reading: 3.0%
To adjust the LRW, the reading needs to be lowered by 0.11%.
3.0% – 0.11% = 2.89%.
This is the new value to be entered into the LRW.
K A M L R W O P E R AT I O N R E A LT E R M C O N T I N U E D
Type the calculated percent of water (i.e. 2.89 for 2.89%) and click on "Send ASCII." Calibration results will be
displayed. FIG. 4-10.
10.
FIG. 4-10
Calibration results