User Manual
OB-22/2 LT
20
4.2.7 Special settings
Default values
Stirring Speed
100%
Temp. Averaging
5
Heat. PowerLimit
100%
4.2.8 Compressor Cooling Settings
Default values
Compr. On under
40°C
Cool.With Hi Pwr
-35
°C
Next screen
Increase value
Decrease value
Touch value to select
Previous screen
Fluid mixing speed
After changing this parameter wait for a few
minutes for the system to stabilize!!!
Never change for more than a few % at a time!
Set the heater power limit
Set the frequency for the temp. Averaging
Next screen
Increase value
Decrease value
Touch value to select
Previous screen
Temperature under which compressor
starts cooling
Temperature to which compressor cools
with full power