Kamoer Fluid Tech(Shanghai) Co.,Ltd. FX-STP WIFI User Manual
18
Calibrate
48.0ml/min
E:120.0rpm
60s
Calibrate
48.0ml/min
120.0rpm
60s
Calibrate
48.0ml/min
120.0rpm
60s
Calibrate
48.0ml/min
120.0rpm
23s
Calibrate
E:48.0ml/min
120.0rpm
60s
Calibrate
49.0ml/min
120.0rpm
60s
a
b
c
d
e
f
Prepare a measuring tube, put the pump pipe out of the pipe into the measuring tube,
into the water, ready to set the standard;
a.
Rotate the button to move the cursor to the speed selection item;
b.
Press the button to enter the speed editing, the speed editor under the spin button
edit the speed, after editing, press the button again to complete the editing;
c.
Rotate the button to the calibration time item and press the button to start the
calibration;
d.
Waiting for the countdown to the calibration to be completed;
e.
When the countdown to the calibration is complete, enter the input calibration result
state, the volume of the liquid in the measuring tube is entered into the interface
by the rotary button, press the button to complete the input, the calibration is
completed.