a.
Flow rate calibration:
click on the flow rate calibration in the setting s --coma
interface, and bring out the flow rate calibration channel selection interface;
b.
Speed setting:
set the speed of calibration;
c.
Start emptying:
the purpose of emptying is to let the air in the pump tube drain,
so that the accuracy of the rear calibration will not be affected;
d.
Long drip timing:
set the running time of the pump during calibration;
Before proceeding to the next step, make sure that the pump pipe inlet has been
immersed in water, the pump pipe outlet is placed in the measuring tube;
e.
Start titration:
click the titration button and the pump will stop after running the
duration of the previous set;
f.
Input volume:
after reading the volume of liquid in the barrel input, in ml;
g.
Calibration complete:
Click the "Calibration complete" button to complete the
flow calibration.