Combo — Manual
3. Calibration Adjustments
On instruments with a push knob, simply push the knob, adjust the
brightness and push the knob once more.
For instruments with only rotation knob, an optional external il-
lumination knob can be connected to the back of the instrument.
Illumination knob part number is I-ALT-ILLUM and it must be or-
dered separately. Please refer also to section 3.3 on page 16 for the
activation procedure.
2.2.5
Power -
POWER
Connect supplied connector at the back of Indu Combo. The connec-
tor has a notch on one side, which protects from wrong orientation.
Connect blue lead to negative (ground) terminal and red lead to
positive (+12-24 V) terminal.
3
Calibration Adjustments
The knob is used to perform some adjustments.
3.1
Altitude Adjustment - Static Sensor Offset
1020
→
999
→
1013 for hPa or 30
.
12
→
29
.
50
→
29
.
91 for inHg.
A minor sensor offset may be required in order to adjust Indu Combo
according to some other reference instrument. Please, follow the steps
given below in order to make the adjustment. A precise reference
instrument is needed in this procedure.
1. Set both the reference instrument and Indu Combo to 1013 hPa
and compare the readings. Write down the altitude shown by
the reference.
14
©
Kanardia
2015-2018