KANE905 Operators Manual
Page 23
PRESS.:
Flue draught can be displayed in millibar (mbar),
hectaPascals (hPa), millimeters water gauge (mmWG) or
inches water gauge (in WG).
EFF.:
Efficiency can be selected for gross or net values. Gross
efficiency assumes latent heat of vaporisation is lost in the
boiler and hence will be lower than net efficiency. For natural
gas the difference will be approximately 11%.
Efficiency is displayed as EFF (G) or EFF (N) respectively.
Should the instrument detect that a condensing boiler is
under test then it automatically switches to a third mode that
is displayed as EFF (C).
SUB MENU -
3. DISPLAY
L
IGHT : OFF
MODE : 8-PAGE
CONTRAST
: DEFAULT
Allows the configuration of the display to be changed.
LIGHT:
Choose from ON or OFF.
MODE:
Select 4 or 8 Page Mode or Line Scroll Mode as detailed in
section MAIN DISPLAYS.
CONTRAST:
The contrast is set to a DEFAULT value or can be adjusted
LIGHTER or
DARKER. Use the
and
keys
to adjust.
SUB MENU -
4. SET UP
The set up menu allows the following parameters to be set / altered.
Language.
Automatic calibration time
CO gas alarm
NOx percentage for calculation
Date and time
Printout header
Printer type