Specifications
Power Supply :
Sensor :
Sensitivity :
Inidcators :
Warm-up :
Response Time :
Duty Cycle :
Battery Life :
Dimensions :
Weight :
Operational Range :
2 x AA alkaline batteries
Solid state semi conductor
<50ppm methane
Audible : Variable tic rate
Visual : Flashing multicoloured LEDs
20 seconds
<1 second
Continuous
5 hours typical use
Handle = 190mm long
Probe = 180 mm long
Approx. 100gm
0-40
o
C , 10% to 90% RH non-condensing
Gases detected include:
Acetone
Alcohol
Ammonia
Benzene
Butane
Ethylene
Gasoline –
Petrol
Hydrogen
Industrial
Solvents
Jet Fuel
Lacquer
Thinners
Methane
Naphtha
Natural Gas
Propane
Toluene
ELECTROMAGNETIC COMPATIBILITY
The European Council Directive 89/336/EEC requires that electronic equipment does
not generate electromagnetic disturbances that exceed defined levels and has an
adequate level of immunity to enable it to be operated as intended. The specific
standards applicable to this product are detailed in the appendices.
Since there are many electrical products in use that pre-date this Directive and may
emit electromagnetic radiation in excess of the standards defined in the Directive
there may be occasions where it would be appropriate to check the analyser prior to
use. The following procedure should be adopted:
•
Go through the normal start up sequence in the location where the equipment is to
be used.
•
Switch on all localised electrical equipment that might be capable of causing
interference.
•
Check that performance is as expected. (A level of disturbance in the
performance is acceptable). If not adjust the position of the instrument to minimise
interference or switch off, if possible, the offending equipment for the duration of
the test.
At the time of writing this manual (December 2006) there is no awareness of any field
based situation where such interference has ever occurred and this advice is only
given to satisfy the requirements of the Directive.