background image

 

KANGAROO E-PUMP INSTRUCTIONS FOR FEED/FLUSH ONLY 

(adapted from Kangaroo ePump Instructions for Use Manual) 

 
 
 

CC.12.10A 

 BC Children’s Hospital Child & Youth Health Policy and Procedure Manual  

Page 1 of 2 

May 1

st

, 2012

                                     Refer to online version – Print copy may not be current – Discard after use

 

 

Battery Life:  

 

o

 

15 hours @125ml per hour 

o

 

8 hour recharge time will recharge while plugged in and running. 

24 Hou Memory:  

o

 

Pump will retain settings and volume to be delivered even if pump is turned off 

Operating Range: 1-400ml in 1ml increments 
E Pump Tubing:  

13ml in tubing

 

 

1. 

ON/OFF: 

o

  To power on, press 

POWER 

button on

 

right hand corner of control panel 

o

  To power off, hold the 

POWER

 button for countdown 3,2,1 until the pump display turns off. Then release 

because if held too long it will turn back on. 

2. 

Keep or Clear Prior Pump Settings:

 Immediately after powering up the pump, if the prior feed or flush settings 

were not cleared, 2 options are given: 

o

 

KEEP SETTINGS 

- to start with the same settings that were most recently programmed into the 

pump. Any of the previously-programmed settings can be modified, if needed. The feed volume 
totals, “

mL FED

” (Continuous Feed Mode) or “

BOLUSES FED

” (Intermittent Feed Mode) and “

mL 

FLUSH

” are also maintained 

o

 

CLEAR SETTINGS 

- to reset all settings and feed totals to zero. It will then be necessary to program 

all settings before running the pump. 

Note:

 The feed totals (“

mL FED

,” “

BOLUSES FED

,” “

mL FLUSH

”) can be cleared at any time, and without 

clearing the pump settings, by doing the following: 

ƒ

  Start the pump 

RUNNING 

ƒ

 Press

HOLD 

ƒ

 Press

CLEAR VOLUME

 

 

ƒ

  the pump will then show 

LOAD A SET 

or, if the Pump Set is already loaded, it will show 

SET 

LOADED 

3. 

Load Feeding Set

:   

a.  Open Blue door, load set. (

Do not overstretch tubing)

 

o

  Use new ePump Feeding Set Only.  

o

  No more Drip Chamber! 

o

 

Hang Height 18 inches: 

in order to achieve proper accuracy, the bottom of the    feeding set 

container must be 18 inches above the top of the feeding pump. 

    

b. Screen 

shows 

CLEAR SETTINGS/KEEP SETTINGS

 (Choose an Option) 

o

  Keep settings = same patient, same prescription 

o

  Clear Setting = resets all settings and feed totals to zero 

 

4. 

Adjust Feed 

o

 

FEED RATE

 (This is the icon to adjust the rate)  

o

  Select Feed Rate and press 

ENTER (

a value must be entered)         

o

 

FEED VTBD 

(Volume to be delivered)

 

o

  Select Volume to be delivered and press

 ENTER (

This is 

optional

 only if you want to use this feature) 

o

  After these have been entered press

 DONE 

to ensure the settings are saved

 

     

5. 

Prime Pump

:  

o

 Press 

AUTO PRIME

 until it shows Autoprime is complete.  

o

 Then 

press 

HOLD TO PRIME FEED

 to complete priming (this will bring the formula to the end of the 

tubing).   

o

 Press 

DONE 

 

6. Press 

RUN

 when ready to start. 

 

7. 

To Stop

, press 

HOLD

 button (This will hold for 10 minutes before it alarms). If you would like to extend the hold, 

press 

RESUME IN __ MIN

 (this will automatically stop the pump for 30 minutes and then it will restart 

automatically). 

Reviews: