10 Spare parts/Accessories 

 

700-0717 

Silicon corner/handle 

Ordered by the piece 

BB-20 

Foam mattress  

For BB-1 and BB-3. Foam mattress, sealed, easy to 
clean 650 x 440 x 40 mm 

BB-21 

Tent for BB-1 

Please check with your local supplier or on the KANMED 
homepage www.KANMED.se 

for colours available 

 

NOTE: Other accessories may be available. Please check with your local supplier or on the KANMED 

homepage 

www.kanmed.se

  

Please contact your supplier for a price list. 

 

 

11 EMC guidelines 

 

Guidance and manufacturer’s declaration – electromagnetic emissions

 

The Babybed is intended for use in the electromagnetic environment specified below. The customer or the user of the Babybed 

should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment - guidance 

RF emissions 

CISPR 11 

Group 1 

The Babybed uses RF energy only for its 

internal function. Therefore, its RF emissions 

are very low and are not likely to cause any 

interference in nearby electronic equipment. 

RF emissions 

CISPR 11 

Class B 

Harmonic emissions 

IEC 61000-3-2 

Class A 

Voltage fluctuations/ flicker emissions 

IEC 61000-3-3 

Complies 

The Babybed is suitable for use in all 

establishments including domestic 

establishments and those directly connected to 

the public low-voltage power supply network 

that supplies buildings used for domestic 

purposes. 

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The Babybed is intended for use in the electromagnetic environment specified below. The customer or the user of the Babybed 

should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level 

Compliance level 

Electromagnetic environment - 

guidance 

Electrostatic discharge 

(ESD) 

IEC 61000-4-2 

+/- 6 kV contact 

+/- 8 kV air 

+/- 6 kV contact 

+/- 8 kV air 

Floors should be wood, concrete or 

ceramic tile.  

If floors are covered with synthetic 

material, 

the relative humidity should be at 

least 30 %. 

Electrical fast transient / 

Burst 

IEC 61000-4-4 

+/- 2 kV for power supply 

lines 

+/- 1 kV for input/output 

lines 

+/- 2 kV for power supply 

lines 

n/a. for input/output lines 

Mains power quality should be that of 

a typical commercial or hospital 

environment. 

Part no 

Description 

Remark 

700-0723 

Side wall left/right 

19 cm height.  

700-0721 

Side wall front 

19 cm height.  

700-0722 

Side wall rear 

19 cm height, Grooves for tubing 

700-0728 

Side wall  left/right 

24 cm height.  

700-0729 

Side wall front 

24 cm height. 

700-0730 

Side wall rear 

24 cm height. Grooves for tubing 

700-0727 Tent 

Pole 

 

 

 

 

Summary of Contents for B-1

Page 1: ...User Manual Art no BB1 070 1 March 2007 Manufactured by KANMED AB www kanmed se Distributed by Caution Please read this manual carefully Incorrect use may cause patient injury ...

Page 2: ...ls 3 3 Preparations before use 4 4 Handling and daily use 5 5 Cleaning and Maintenance 6 6 Trouble shooting 6 7 Warranty Statement 6 8 Technical Data 7 9 Technical service and maintenance 8 10 Spare parts Accessories 9 11 EMC guidelines 9 KANMED BABYBED TWIN VERSION AND FIXED HEIGHT VERSION ...

Page 3: ... weight limits Do not place electronic equipment on the shelf under the bed 2 Description of Symbols Maximum Load permitted Bed surface for BB 1 and BB 3 10 kg Excl BABYWARMER Control Unit 3kg and Water mattress 4 5kg Bed surface BB 4 10 kg Excl Two BABYWARMER Control Unit 2x3kg and two Water mattresses 2x4 5kg or one BABYWARMER Control Unit 3kg and twin water mattress 9kg Shelf under bed surface ...

Page 4: ...ent has a channel inside Thread the Tent pole into the channel in order to secure it Place the mattress in the bed 3 4 Check all bolts and screws Make sure that all bolts and screws are properly tightened and that the bed feels stable Run the electrical bed up and down and ensure that it moves smoothly without any unusual noise 3 5 Kanmed Babywarmer The Kanmed Babywarmer control unit is hung on it...

Page 5: ... for BB 3 The height is adjustable by the pushbuttons on the front The lifting column has an overload protection that will not be activated under normal conditions Constantly adjusting height with heavy loads may activate the overload protection When the motor has cooled down the lifting column works as normal again Warning Do not block the movements when activating the electrical height adjustmen...

Page 6: ...s not keep the position Something is blocking it Check that nothing blocks it Bed feels unstable Loose screws or wheels Technical service Height adjustments makes unusual noises Technical service 7 Warranty Statement KANMED warrants the purchaser that the Babybed is free from defects in materials and workmanship for a period of 12 month from the date of delivery The sole obligation of KANMED with ...

Page 7: ...s I Class I NA Used standards SS EN 60 601 SS EN 60 601 Products carries the CE mark Conforms to UL 60601 1 Certified to CAN CSA C22 1 No 601 1 EMC The Kanmed Babybed contains no electronics sensitive to EMC Highly sensitive equipment connected to the same electrical outlet as the Kanmed Babybed may be affected by the lifting column electrical motor The motor may generate a moderate amount of flic...

Page 8: ...pring from the pillar Start by taking away the small safety spring Make someone or something hold the bed Pull out the gas spring from the pillar Un tighten the locking nut by turning the gas spring you adjust the contact with the release arm 9 3 Changing side walls The easiest way to replace a side wall is to place it in its hanging down position insert one pin in the lower grove bend the wall an...

Page 9: ... 3 Complies The Babybed is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity The Babybed is intended for use in the electromagnetic environment specified below The customer or the user ...

Page 10: ...ltage prior to application of the test level Guidance and manufacturer s declaration electromagnetic immunity The Babybed is intended for use in the electromagnetic environment specified below The customer or the user of the Babybed should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF co...

Page 11: ...ed The Babybed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Babybed can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the Babybed as recommended below according to the maximum output power of the communicati...

Page 12: ...KANMED AB Gårdsfogdevägen 18B SE 168 66 BROMMA Stockholm Sweden Telephone 46 8 56 48 06 30 Telefax 46 8 56 48 06 39 E Mail info kanmed se Home page www kanmed se Distributed by ...

Reviews: