User Manual BB1/070/1
Page 7 of 12
8 Technical Data
BB-1 Standard
BB-4 Twin
BB-3 Fixed height
Outer
dimensions:
760mm x 560mm 850x700mm
760mm x 560mm
Bed surface:
Inner dims
665mm x 445mm
Inner dims
765mmX640mm
Inner dims
665mm x 445mm
Height:
Max 990mm
Max 990mm
Fixed to 800mm
Min 700mm
Min 700mm
Weight:
About 33 kg
About 40kg
About 25 kg
Wheels base:
700mm x 500mm
700mm x 500mm
700mm x 500mm
Wheels:
Diameter 125mm.
Diameter 125mm
Diameter 125mm.
Front wheels with brakes.
Front wheels with brakes.
Front wheels with
brakes.
Roller bearings.
Roller bearings
Roller bearings.
Tilt:
± 22°
± 22°
± 22°
Mattress:
Polyurethane with PVC
cover. (665x445x30mm)
Polyurethane with PVC
cover. (765x640x30mm)
Polyurethane with PVC
cover. (665x445x30mm)
Tent:
100% Bedton. Washable
60°C
100% Bedton. Washable
60°C
100% Bedton. Washable
60°C
Shelf:
Min 400mm x 400mm
Min 400mm x 400mm
Min 400mm x 400mm
Colour:
Unisex light blue
Unisex light blue
Unisex light blue
Voltage
100-240Vac, 50-60Hz
100-240Vac, 50-60Hz
NA
Power: 127VA
127VA
NA
Fuses: NA
NA
Protection:
Class I
Class I
NA
Used standards:
SS EN 60 601
SS EN 60 601
Products carries the CE-mark.
Conforms to UL 60601-1 Certified to CAN / CSA – C22.1 No 601.1
EMC
The Kanmed Babybed contains no electronics sensitive to EMC.
Highly sensitive equipment connected to the same electrical outlet as the Kanmed Babybed may be
affected by the lifting column electrical motor. The motor may generate a moderate amount of flicker
during start.
Environmental data
In use: +10 to + 40ºC, <85% RH
Transport: -20 to +60ºC, non condensing.
Lifting column limitations
Observe that the total load in the bed surface is limited to 10kg (BABYWARMER system and
water mattress not included).
The electrical motor in the lifting column is not designed for continues use.
Continued operation should be limited to 2 min, max 5 cycles per minute.
Modifications
Any modifications on Kanmed Babybed including home made accessories will
void KANMED’s responsibilities totally and are not allowed without written
consent of KANMED.
Creating Systems
Any person connecting the Kanmed Babybed to any other systems mains
socket may have created a “system” according to IEC 60601-1 section 16,
and must be qualified to evaluate any consequences that may harm staff,
patient and equipment.