Missing Water Mattress alarm   H

2

1.

 

Start the unit and observe an “OK” self test. 

2.

 

Place the Heating Pad without the Water Mattress on a flat surface. 

3.

 

Wait 5 to 7 minutes and check that the H2O alarm lamp is lit. 

NOTE:  The Heating pad must be at room temperature before starting this test. 

 

Heating Pad fault alarm 

For this test the heating pad is not connected to the control unit. 
Start the warming and check that the Heating Pad symbol and the Pad lamp are lit after about 3 sec. and 
that the beeper sounds

 

Supply voltage failure alarm 

For this test the unit must be assembled completely with Water Mattress and Heating Pad 

1.

 

Start the warming  

2.

 

After about one minute, select a temperature different from 37

 o

C and make a note of it. 

3.

 

Wait about 2 minutes and then disconnect the supply voltage by using the mains switch or pulling 
out the mains plug.  

4.

 

Check that the temperature lamp is off and that the Control unit lamp starts to blink and that the 
acoustic alarm beeps intermittently for at least 10 minutes 

5.

 

Switch on supply voltage again and check that the self test is performed OK and that the warming 
starts at the value selected before supply voltage interruption. 

 

Internal 1.5 V battery test 

This test may only be done by a qualified technician. During this test the heating pad must be connected 
and the internal 1,5 Volt battery removed from its holder by opening the Control Unit. 
 

1.

 

Press start and check that the red lamps for, 

T

H

2

O

 and              blinks. Press Start/standby 

2.

 

Make a new start but keep the alarm button pressed continuously until the self test has ended. 

3.

 

Check that the warming starts i.e. that the missing or "dead" battery can be bypassed. 

 

11   Accessories, Spare parts and technical documentation 

 

Your local supplier has all information about available spare parts and accessories or check 

www.KANMED.se 

User Manuals in many languages can be downloaded from www.KANMED.se

 

 

KANMED or your local distributor will provide a spare part price list as well as a service manual that 

contains all information necessary to perform maintenance, and simpler repairs.

 

 

Article number  

Description 

Quantity 

BW-50-002 

Heating Pad BW50 

BW-50-003 

Water Mattress BW50 

BW-50-025 

KANMED Baby Nest 

1( 5 pack available) 

BW-50-005 

Filling Tube BW50 

BW-50-029 

Anti algae 

Box of 20 

BW-50-007 

Battery Cable  for Control Unit 
BW50 

Spare parts  

 

 

700-0001 

Fuse T 1A ( F1,F2, 220 -240V) 

Pack of 10 

700-0079 

Fuse T 2A (F1,F2, 110 -115 V) 

Pack of 10 

700-0324 

Fuse  T 3,15A (F3 + F4) 

Pack of 10 

699-1171 

Mains Cord 230V 

 

 

Summary of Contents for BABYWARMER

Page 1: ...User Manual Art No BW 50 070 10 2007 03 Caution Please read this manual carefully Incorrect use of patient warming equipment may cause patient injury Manufactured by KANMED AB Sweden Distributed by 0413 ...

Page 2: ...n 11 9 Trouble shooting 11 10 Safety systems available 12 11 Accessories Spare parts and technical documentation 13 12 Technical data 14 13 Warranty 15 14 Disposal 15 NOTE This manual is valid for all KANMED Baby Warmer of model KBW50 recognised by the blue silicone rubber edges on the control unit This user manual contains important safety information and must be read carefully and shall be saved...

Page 3: ...2 Connect the power cable at the back of the Control Unit and switch it on 3 Hang the Control Unit in a safe place where it can be seen 4 Press the start button on the front panel and observe that it starts at 370 C after a short self test Placing the baby 1 Only when the desired temperature is reached you can place the baby in the nest 2 Position the lightly dressed baby diaper long arm T shirt b...

Page 4: ...led Temp lock in text Alarm button To silence T alarm for 10 minutes Over temperature alarm T Indicates water temperature deviation more than 1 o C from selected value H2O Indicates that the Water Mattress is missing or in the wrong position Indicates Heating Pad fault Certified by SEMKO according to EN 60 601 1 EN 60 601 1 2 EN 60 601 2 35 CE0413 Fullfills MDD 93 42 EEC EC Type examination certif...

Page 5: ...e of the Heating Pad is turned towards the Water Mattress Make sure that there are at least 2 holes in the bottom of the bed If feasible the connection cable to the Heating Pad can be thread through a suitably placed hole Always place the baby with its head towards the closed semicircular part of the KANMED Baby Nest Make sure that the adjusting string is placed outside the Baby Nest to prevent it...

Page 6: ...ad The design and construction of the Heating Pad has reduced magnetic and electrical fields to normal background levels and thereby making them harmless The voltage to the Heating Pad is SELV 24V DC Safe Extra Low Voltage See technical data for information about warming up times Explanation of the drawings on the Heating Pad Please make them available while reading this 1 Heating Pad BW 50 002 ma...

Page 7: ... Air pressure may build while filling Interrupt filling and release air if necessary 11 Remove air bubbles Very Important Place the Water Mattress on a flat surface and lift the filling hole 5 to 10 cm upwards Press the air bubbles towards the hole repeatedly until all major air bubbles are expelled 12 Lower the filling hole while pressing the Water Mattress so that no air enters and insert the pl...

Page 8: ... Only for indication Alarms Description and checks of the function of alarms is found in the technical section Stop the acoustic alarm by pressing the alarm button or switch the unit off by pressing the Start Stand By button Only T can be cancelled for 10 minutes by pressing the alarm button Try to rectify the cause for an alarm see section trouble shooting and switch on the unit again If the alar...

Page 9: ...ion of the KANMED Baby Warmer is not influenced by the incubator Try it for children in need of a soft warm support for instance children suffering from hydrocephalus fractures or headaches after delivery with a forceps or a suction device colicky pains etc NIDCAP Developmental Care and Kangaroo care In these treatment forms the sleeping and rest as well as positioning of the baby plays an essenti...

Page 10: ...illing material must be changed as soon as it shows signs of damage or wear at least after one year of use Periodical Checks Recommended checks To be done by Interval Water Mattress and water level Heating Pad Baby Nest are they clean and undamaged The user Every time the bed is made Cables and connectors are they clean and undamaged The user At each start Self test at start The user At each start...

Page 11: ... stops Supply voltage failure Supply voltage mains or battery has disappeared or been switched off without pressing Start Standby first Stops Internal battery T H2O Internal battery is weak or missing at start up Press and hold the Alarm button when restarting Replace battery as soon as possible Will not start 9 Trouble shooting Symptom Probable cause Remedy No supply voltage connected Connect sup...

Page 12: ...amps blinks 5 times indicating PROM version 4 Temperature lamp 37 is lit 5 If the self test is performed as described above it is ready for use Over temperature alarm 1 Start the unit and observe self test is OK 2 Fill a bucket Height about 30 cm and min diameter 20 cm with warm water 38 o C 3 Roll the Heating Pad loosely and insert it fully in the water 4 Slowly fill with warm water until the tem...

Page 13: ...alue selected before supply voltage interruption Internal 1 5 V battery test This test may only be done by a qualified technician During this test the heating pad must be connected and the internal 1 5 Volt battery removed from its holder by opening the Control Unit 1 Press start and check that the red lamps for T H2O and blinks Press Start standby 2 Make a new start but keep the alarm button pres...

Page 14: ...nstalled according to section 2 Internal Battery Alkaline 1 5V size LR6 AA Only for power failure alarm Water Mattress Dimensions and Weight Length 600 mm With 270 mm Height 30 mm Other sizes might be available Please contact your local supplier Weight about 0 5 kg Material PVC Anti Algae recommended KANMED Bottle 119 ml fluid Poly Oxyethylene dimethylimino ethylene dimethyliminio ethelyenedichlor...

Page 15: ...d KANMED s responsibilities totally and are not allowed without the written consent of KANMED 13 Warranty KANMED warrants the purchaser that the Control Unit BW 50 Heating Pad BW 50 002 are free from defects in material and workmanship for a period of 12 month from the date of delivery The KANMED Baby Nest and associated parts are warranted to be free of defects at the time of delivery The sole ob...

Page 16: ...nual BW 50 070 10 Page 16 of 16 KANMED AB Gårdsfogdevägen 18B SE 168 66 BROMMA Stockholm Sweden Telephone 46 8 56 48 06 30 Telefax 46 8 56 48 06 39 E Mail info KANMED se Home page www KANMED se Distributed by ...

Reviews: