background image

 

 

                                        

User Manual 

BW-50-070/10

 

 

 

Page 7 of 16  

The Water Mattress 

When filled to the level mark it contains about 4.5 litres of water. If you plan to tilt the bed then add 2 cm 

of water. Fill by using the filling tube. Add Anti Algae every 3 months. Water molecules diffuse through the 
material. This is normal. Therefore the water level should be checked regularly 

NOTE: Do not tilt the Water Mattress if the temperature must be higher than 37

O

C. The very 

sensitive over temperature circuit can alarm and switch off the warming. 
When tilting during warming up, too little water will make the over temperature alarm go off. 

Explanation of the drawings on the Water Mattress  

 

9.

 

The level mark on the Water Mattress indicates standard water amount. If tilting is required please 

add 2 cm more water.

 

10.

 

Place the Water Mattress in an upright position and fill it with the enclosed Filling Tube. When the 

desired level has been reached, add Anti Algae carefully through the filling hole or via the Filling 

Tube. Follow the caution instruction on the Anti Algae package. 

NOTE: Air pressure may build 

while filling. Interrupt filling and release air if necessary.

 

 

11.

 

Remove air bubbles, Very Important. Place the Water Mattress on a flat surface and lift the filling 

hole 5 to 10 cm upwards. Press the air bubbles towards the hole repeatedly until all major air 
bubbles are expelled. 

12.

 

Lower the filling hole while pressing the Water Mattress so that no air enters and insert the plug. 

13.

 

Turn the Water Mattress upside down and insert the Heating Pad in the pocket.  

14.

 

The Water Mattress and the Heating Pad is inserted into the KANMED Baby Nest. 

 

 

The KANMED Baby Nest 

Different types of KANMED Baby Nests are available. Please check with your supplier or on internet 

www.KANMED.se 
 

The KANMED Baby Nest is intended to prevent the baby from rolling off the water mattress, to facilitate 

correct positioning and to create a tight and cosy environment around the baby. 

By pulling the strings one can adjust the collar from being wide open and flat to become a tight raised wall 
around the baby. The Baby Nest consists of a bag into which the Water Mattress and Heating Pad is 

inserted.  

Note:

 Check the Baby Nest regularly for damages. To reduce the risk of a nest splitting during washing we 

recommend that the nest is used for only one year. 

 

 
 

 ! 

Summary of Contents for BABYWARMER

Page 1: ...User Manual Art No BW 50 070 10 2007 03 Caution Please read this manual carefully Incorrect use of patient warming equipment may cause patient injury Manufactured by KANMED AB Sweden Distributed by 0413 ...

Page 2: ...n 11 9 Trouble shooting 11 10 Safety systems available 12 11 Accessories Spare parts and technical documentation 13 12 Technical data 14 13 Warranty 15 14 Disposal 15 NOTE This manual is valid for all KANMED Baby Warmer of model KBW50 recognised by the blue silicone rubber edges on the control unit This user manual contains important safety information and must be read carefully and shall be saved...

Page 3: ...2 Connect the power cable at the back of the Control Unit and switch it on 3 Hang the Control Unit in a safe place where it can be seen 4 Press the start button on the front panel and observe that it starts at 370 C after a short self test Placing the baby 1 Only when the desired temperature is reached you can place the baby in the nest 2 Position the lightly dressed baby diaper long arm T shirt b...

Page 4: ...led Temp lock in text Alarm button To silence T alarm for 10 minutes Over temperature alarm T Indicates water temperature deviation more than 1 o C from selected value H2O Indicates that the Water Mattress is missing or in the wrong position Indicates Heating Pad fault Certified by SEMKO according to EN 60 601 1 EN 60 601 1 2 EN 60 601 2 35 CE0413 Fullfills MDD 93 42 EEC EC Type examination certif...

Page 5: ...e of the Heating Pad is turned towards the Water Mattress Make sure that there are at least 2 holes in the bottom of the bed If feasible the connection cable to the Heating Pad can be thread through a suitably placed hole Always place the baby with its head towards the closed semicircular part of the KANMED Baby Nest Make sure that the adjusting string is placed outside the Baby Nest to prevent it...

Page 6: ...ad The design and construction of the Heating Pad has reduced magnetic and electrical fields to normal background levels and thereby making them harmless The voltage to the Heating Pad is SELV 24V DC Safe Extra Low Voltage See technical data for information about warming up times Explanation of the drawings on the Heating Pad Please make them available while reading this 1 Heating Pad BW 50 002 ma...

Page 7: ... Air pressure may build while filling Interrupt filling and release air if necessary 11 Remove air bubbles Very Important Place the Water Mattress on a flat surface and lift the filling hole 5 to 10 cm upwards Press the air bubbles towards the hole repeatedly until all major air bubbles are expelled 12 Lower the filling hole while pressing the Water Mattress so that no air enters and insert the pl...

Page 8: ... Only for indication Alarms Description and checks of the function of alarms is found in the technical section Stop the acoustic alarm by pressing the alarm button or switch the unit off by pressing the Start Stand By button Only T can be cancelled for 10 minutes by pressing the alarm button Try to rectify the cause for an alarm see section trouble shooting and switch on the unit again If the alar...

Page 9: ...ion of the KANMED Baby Warmer is not influenced by the incubator Try it for children in need of a soft warm support for instance children suffering from hydrocephalus fractures or headaches after delivery with a forceps or a suction device colicky pains etc NIDCAP Developmental Care and Kangaroo care In these treatment forms the sleeping and rest as well as positioning of the baby plays an essenti...

Page 10: ...illing material must be changed as soon as it shows signs of damage or wear at least after one year of use Periodical Checks Recommended checks To be done by Interval Water Mattress and water level Heating Pad Baby Nest are they clean and undamaged The user Every time the bed is made Cables and connectors are they clean and undamaged The user At each start Self test at start The user At each start...

Page 11: ... stops Supply voltage failure Supply voltage mains or battery has disappeared or been switched off without pressing Start Standby first Stops Internal battery T H2O Internal battery is weak or missing at start up Press and hold the Alarm button when restarting Replace battery as soon as possible Will not start 9 Trouble shooting Symptom Probable cause Remedy No supply voltage connected Connect sup...

Page 12: ...amps blinks 5 times indicating PROM version 4 Temperature lamp 37 is lit 5 If the self test is performed as described above it is ready for use Over temperature alarm 1 Start the unit and observe self test is OK 2 Fill a bucket Height about 30 cm and min diameter 20 cm with warm water 38 o C 3 Roll the Heating Pad loosely and insert it fully in the water 4 Slowly fill with warm water until the tem...

Page 13: ...alue selected before supply voltage interruption Internal 1 5 V battery test This test may only be done by a qualified technician During this test the heating pad must be connected and the internal 1 5 Volt battery removed from its holder by opening the Control Unit 1 Press start and check that the red lamps for T H2O and blinks Press Start standby 2 Make a new start but keep the alarm button pres...

Page 14: ...nstalled according to section 2 Internal Battery Alkaline 1 5V size LR6 AA Only for power failure alarm Water Mattress Dimensions and Weight Length 600 mm With 270 mm Height 30 mm Other sizes might be available Please contact your local supplier Weight about 0 5 kg Material PVC Anti Algae recommended KANMED Bottle 119 ml fluid Poly Oxyethylene dimethylimino ethylene dimethyliminio ethelyenedichlor...

Page 15: ...d KANMED s responsibilities totally and are not allowed without the written consent of KANMED 13 Warranty KANMED warrants the purchaser that the Control Unit BW 50 Heating Pad BW 50 002 are free from defects in material and workmanship for a period of 12 month from the date of delivery The KANMED Baby Nest and associated parts are warranted to be free of defects at the time of delivery The sole ob...

Page 16: ...nual BW 50 070 10 Page 16 of 16 KANMED AB Gårdsfogdevägen 18B SE 168 66 BROMMA Stockholm Sweden Telephone 46 8 56 48 06 30 Telefax 46 8 56 48 06 39 E Mail info KANMED se Home page www KANMED se Distributed by ...

Reviews: