background image

USER and TECHNICAL MANUAL KANMED Warming Cabinet GE-2-1370     

Page 13 of 16 

12 

CIRCUIT DIAGRAM for cabinets delivered from 2004 

  
 

 
 

  

1.  

Connection socket 

2.   F1 

Fuse  6,3A 

 

 

4.   S2 

Main switch  

5.   T1 

Temperature regulator 

6.   T2 

Over temperature protection 

7.   L1 

Warning lamp red – over temperature warning 

8.   M1 

Fan 

9.   G1 

Temperature sensor 

10. TR 

Transformer 

11. RE 

Relays 

12. HE 

Heating element 

13. F2 

Fuses 50mA 

Summary of Contents for GE-1380

Page 1: ...ns important information concerning safety and daily use as well as maintenance instructions and should be stored for future use This manual is valid for all cabinets that are delivered after 2001 Manufactured by KANMED AB January 2004 PART NO GE 1002 070 3 ...

Page 2: ...DESCRIPTION 4 3 DESCRIPTION OF FUNCTION 4 4 TECHNICAL DATA 5 5 UNPACKING AND PACKING LIST 6 6 INSTALLATION 6 7 USING THE CABINET 6 8 MAINTENANCE AND TROUBLE SHOOTING 8 9 ACCESSORIES AND SPARE PARTS 10 10 WARRANTY 11 11 CIRCUIT DIAGRAM FOR CABINETS DELIVERED UNTIL 2003 12 12 CIRCUIT DIAGRAM FOR CABINETS DELIVERED FROM 2004 13 ...

Page 3: ...skets at the same time The cabinet can tilt forward if it is not secured to the wall Take extra caution if the cabinet is mounted on wheels especially when the cabinet is moved around Note If the cabinet is not secured to a wall don t pull out more than one shelf basket at a time Do not overload the shelves basket maximum 8 kilos per shelf Max 2 pcs of gel pads 45x50x1 5 Max load of the basket is ...

Page 4: ...s are available Mount the rails with the screws supplied using the prepared mounting holes and then push the shelf basket in place by holding the front higher than the rear 3 DESCRIPTION OF FUNCTION The warming cabinet and its contents are warmed by warm air heated by a 1000 Watt warming element The hot air is circulated by a fan and distributed evenly through the outlets at the rear of the cabine...

Page 5: ...r measure Height 13cm Width 52cm Dept 50cm Maximum load 20kg TEMPERATURES Working temperature 35 C 42 C selectable in steps of 1 C Of circulating air Over temperature protection Visual alarm 45 C Accuracy 1 C Hysteresis T2 overtemp 4 C max WARMING CAPACITY Warming of 10 gel pads GE 455015 3 hours 45 x 50 x 1 5 cm from 22 C to 40 MODE OF OPERATION Designed for continuous use CE MARKING According to...

Page 6: ...NOTE Always secure the cabinets to a wall by using the built in top brackets to avoid the risk of tipping forward at unfortunate circumstances 7 USING THE CABINET Open the door and switch on the Cabinet with the green power switch at the upper right side Check the set temperature and adjust if necessary according to 7 2 7 1 GENERAL ADVICE The Cabinet is designed for continuous use and shall be lef...

Page 7: ...OWN SET Press set button to show the set selected temperature The set value is shown and the out lamp blinks as long as the set button is pressed While the set button is pressed the set value can be altered by pressing the UP DOWN button UP Press for temperature rise DOWN Press for temperature decrease OUT Flashes as long as the set button is pressed Lights permanently while heating takes place re...

Page 8: ...scription below NOTE To adjust the temperature T1 you have to have the door open On older cabinets equipped with a door switch the door switch S1 must be pressed with for example some tape while you do the correction Don t forget to remove the tape when you are finished PROCEDURE Place the external control sensor as described under GENERAL INFORMATION above Start the warming and wait until the tem...

Page 9: ...ON ABOUT TEMPERATURE MEASUREMENT AND CALIBRATION The temperature sensor T2 is a capillary thermostat In the KANMED cabinet its hysteretic is 4 C max When delivered T2 is set to 2 C above the maximum value of T1 42 C The activation of T2 is identified by a clear click sound as well as by the red over temperature lamp is lit The procedure described below is a simplified method to check and if necess...

Page 10: ...ut the element frequently there is not enough airflow through the element Check the fan and that the air intake under the heating compartment is not blocked NOTE All checking that involves opening the heating compartment must be done by a qualified technician 9 ACCESSORIES AND SPARE PARTS ACCESSORIES Part no Description Quantity GE 41500 Shelf 1 GE 41600 Basket 1 SPARE PARTS Part no Pos in drawing...

Page 11: ...t apply if the product has been modified adjusted or repaired other than by KANMED or by organisations authorised by KANMED or modified adjusted or repaired not in accordance with written instructions provided by KANMED or if the equipment has been subject to misuse negligence or accident These warranties are made on the condition that prompt notification of a defect is given to KANMED or its auth...

Page 12: ...red until 2003 1 Connection socket 2 F1 Fuse 6 3 A 3 S1 Safety switch activated by the door 4 S2 Mains switch 5 T1 Temperature regulator 6 T2 Over temperature protection 7 L1 Warning lamp red over temperature warning 8 M1 Fan 9 G1 Temperature sensor 10 TR Transformer 11 RE Relays 12 HE Heating element 13 F2 Fuse ...

Page 13: ...for cabinets delivered from 2004 1 Connection socket 2 F1 Fuse 6 3A 4 S2 Main switch 5 T1 Temperature regulator 6 T2 Over temperature protection 7 L1 Warning lamp red over temperature warning 8 M1 Fan 9 G1 Temperature sensor 10 TR Transformer 11 RE Relays 12 HE Heating element 13 F2 Fuses 50mA ...

Page 14: ...USER and TECHNICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 14 of 16 ...

Page 15: ...USER and TECHNICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 15 of 16 ...

Page 16: ...ICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 16 of 16 KanMed AB Gårdsfogdevägen 18B S 168 66 BROMMA SWEDEN Tel 46 0 8 564 80 630 Fax 46 0 8 564 80 639 E Mail info KANMED se Home page www KANMED se DISTRIBUTED BY ...