USER and TECHNICAL MANUAL KANMED Warming Cabinet GE-2-1370
Page 4 of 16
2 GENERAL
DESCRIPTION
THE CABINET
The cabinet is made of stainless steel and insulated in order to reduce heat loss to ambience
and to reduce noise. For the same reason the door is double glassed.
THE HEATING COMPARTMENT
Heating element, fan, thermostat, temperature regulator/indicator and electrical connections
are all placed on a "shelf" at the top of the cabinet. A spare fan is also included and mounted
in the heating compartment. The heating compartment is kept in place by a small screw that,
when loosened, allows the whole heating compartment to be pulled completely out to make
service quick and easy.
THE SHELVES-BASKETS
The shelves/baskets run on wheels and can be pulled fully out until they are automatically
stopped. To completely remove the shelves/basket for cleaning or repositioning, lift the front
of about 5 cm upwards and remove it. The shelves can carry maximum 8 kilos, basket 20 kg.
EXTRA SHELVES- EXTRA BASKETS
Extra shelves/baskets and rails are available.
Mount the rails with the screws supplied using the prepared mounting holes and then push
the shelf/basket in place by holding the front higher than the rear.
3
DESCRIPTION OF FUNCTION
The warming cabinet and its contents are warmed by warm air heated by a 1000 Watt
warming element. The hot air is circulated by a fan and distributed evenly through the outlets
at the rear of the cabinet. The temperature is regulated, by the temperature regulator T1 to
the set temperature.
Thermostat T2 functions as an over temperature protection and will take over the
temperature control in case the air temperature exceeds 45
°
C. At the same time the red
lamp on the front panel will be lit to indicate that there is a malfunction. Inside the heating
element there is an additional over temperature protection that is self-resetting. It will be
activated in case the fan stops or goes too slow.
On old versions of the cabinets a power switch at the door interrupts the power when the
door is opened. When restarted again the temperature goes back to the set temp as it was
before interrupted.
On the new cabinets this power switch is removed
NOTE: If the red over temperature lamp is lit there is an error that requires a technician.