8 MAINTENANCE 

AND 

TROUBLE SHOOTING 

8.1 CLEANING 

Clean and disinfect with normal detergents. If hepatitis or HIV is suspected then use stronger 
disinfectants. 

8.2 TEMPERATURE 

CONTROL/CALIBRATION 

To be performed by a qualified technician only.

 

 
GENERAL INFORMATION

 

The working temperature is regulated by the temperature regulator T1 and the over 
temperature protection by the sensors T2. When control, calibration of T1 and T2, their value 
is compared to the value of an external precision thermometer. ( for example the KANMED 
article Baywatch-01) 
The sensor T2 can be adjusted through a hole in the bottom of the heating compartment. 
The external control thermometer, with a precision temperature sensor, is to be placed 
between two gel pads (GE-455015 or GE-255015) in the middle of the centre shelf. 
Let the cabinet warm up properly (at least 12 hours if cold start) before attempting calibration. 
  

8.2.1 TEMPERATURE REGULATION 

 
INFORMATION ABOUT TEMPERATURE MEASUREMENT AND CALIBRATION

 

T1 is a powered microprocessor temperature regulator/indicator that has resolution of 1

°

C. 

Corrections can be done through a series of pushing’s on the regulator buttons according to 
the description below. 
 

NOTE! To adjust the temperature T1 you have to have the door open. On older 
cabinets equipped with a door switch the door switch S1 must be pressed with for 
example some tape while you do the correction. Don’t forget to remove the tape when 
you are finished.  

 

PROCEDURE 

Place the external control sensor as described under GENERAL INFORMATION above. 
Start the warming and wait until the temperature is stable (at least 12 hours). Compare the 
external control thermometer with the set temperature and if the deviation is bigger than +/-
1

°

C adjust as follows. 

 
Old regulator type (700-0464) 

1.  Press UP and DOWN exactly simultaneously (only one beep may be heard, if you 

don’t succeed- .repeat) and keep pressed for 5 sec until the indicator display shows 
“PA” 

2.  Press UP or DOWN until the display shows “1” 
3.  Press  SET and at the same time UP or DOWN to compensate the noted temperature 

difference, one press gives a adjustment of 1

°

C.  

1.  For example: The display shows 42

°

C but the external control temperature shows 

39

°

C. The difference is -3

°

C, to compensate this difference press three times on 

DOWN button (the earlier value will drop with 3 degrees) 

4.  Store the new calibrated value by pressing UP and DOWN simultaneously (only one 

beep may be heard, if you don’t succeed- .repeat) 

5.  Measure the temperature again and control that T1: s shown value compares to that 

of the external thermometer. 

  

If necessary dismount the locking of the door switch S1 
 

Summary of Contents for GE-1380

Page 1: ...ns important information concerning safety and daily use as well as maintenance instructions and should be stored for future use This manual is valid for all cabinets that are delivered after 2001 Manufactured by KANMED AB January 2004 PART NO GE 1002 070 3 ...

Page 2: ...DESCRIPTION 4 3 DESCRIPTION OF FUNCTION 4 4 TECHNICAL DATA 5 5 UNPACKING AND PACKING LIST 6 6 INSTALLATION 6 7 USING THE CABINET 6 8 MAINTENANCE AND TROUBLE SHOOTING 8 9 ACCESSORIES AND SPARE PARTS 10 10 WARRANTY 11 11 CIRCUIT DIAGRAM FOR CABINETS DELIVERED UNTIL 2003 12 12 CIRCUIT DIAGRAM FOR CABINETS DELIVERED FROM 2004 13 ...

Page 3: ...skets at the same time The cabinet can tilt forward if it is not secured to the wall Take extra caution if the cabinet is mounted on wheels especially when the cabinet is moved around Note If the cabinet is not secured to a wall don t pull out more than one shelf basket at a time Do not overload the shelves basket maximum 8 kilos per shelf Max 2 pcs of gel pads 45x50x1 5 Max load of the basket is ...

Page 4: ...s are available Mount the rails with the screws supplied using the prepared mounting holes and then push the shelf basket in place by holding the front higher than the rear 3 DESCRIPTION OF FUNCTION The warming cabinet and its contents are warmed by warm air heated by a 1000 Watt warming element The hot air is circulated by a fan and distributed evenly through the outlets at the rear of the cabine...

Page 5: ...r measure Height 13cm Width 52cm Dept 50cm Maximum load 20kg TEMPERATURES Working temperature 35 C 42 C selectable in steps of 1 C Of circulating air Over temperature protection Visual alarm 45 C Accuracy 1 C Hysteresis T2 overtemp 4 C max WARMING CAPACITY Warming of 10 gel pads GE 455015 3 hours 45 x 50 x 1 5 cm from 22 C to 40 MODE OF OPERATION Designed for continuous use CE MARKING According to...

Page 6: ...NOTE Always secure the cabinets to a wall by using the built in top brackets to avoid the risk of tipping forward at unfortunate circumstances 7 USING THE CABINET Open the door and switch on the Cabinet with the green power switch at the upper right side Check the set temperature and adjust if necessary according to 7 2 7 1 GENERAL ADVICE The Cabinet is designed for continuous use and shall be lef...

Page 7: ...OWN SET Press set button to show the set selected temperature The set value is shown and the out lamp blinks as long as the set button is pressed While the set button is pressed the set value can be altered by pressing the UP DOWN button UP Press for temperature rise DOWN Press for temperature decrease OUT Flashes as long as the set button is pressed Lights permanently while heating takes place re...

Page 8: ...scription below NOTE To adjust the temperature T1 you have to have the door open On older cabinets equipped with a door switch the door switch S1 must be pressed with for example some tape while you do the correction Don t forget to remove the tape when you are finished PROCEDURE Place the external control sensor as described under GENERAL INFORMATION above Start the warming and wait until the tem...

Page 9: ...ON ABOUT TEMPERATURE MEASUREMENT AND CALIBRATION The temperature sensor T2 is a capillary thermostat In the KANMED cabinet its hysteretic is 4 C max When delivered T2 is set to 2 C above the maximum value of T1 42 C The activation of T2 is identified by a clear click sound as well as by the red over temperature lamp is lit The procedure described below is a simplified method to check and if necess...

Page 10: ...ut the element frequently there is not enough airflow through the element Check the fan and that the air intake under the heating compartment is not blocked NOTE All checking that involves opening the heating compartment must be done by a qualified technician 9 ACCESSORIES AND SPARE PARTS ACCESSORIES Part no Description Quantity GE 41500 Shelf 1 GE 41600 Basket 1 SPARE PARTS Part no Pos in drawing...

Page 11: ...t apply if the product has been modified adjusted or repaired other than by KANMED or by organisations authorised by KANMED or modified adjusted or repaired not in accordance with written instructions provided by KANMED or if the equipment has been subject to misuse negligence or accident These warranties are made on the condition that prompt notification of a defect is given to KANMED or its auth...

Page 12: ...red until 2003 1 Connection socket 2 F1 Fuse 6 3 A 3 S1 Safety switch activated by the door 4 S2 Mains switch 5 T1 Temperature regulator 6 T2 Over temperature protection 7 L1 Warning lamp red over temperature warning 8 M1 Fan 9 G1 Temperature sensor 10 TR Transformer 11 RE Relays 12 HE Heating element 13 F2 Fuse ...

Page 13: ...for cabinets delivered from 2004 1 Connection socket 2 F1 Fuse 6 3A 4 S2 Main switch 5 T1 Temperature regulator 6 T2 Over temperature protection 7 L1 Warning lamp red over temperature warning 8 M1 Fan 9 G1 Temperature sensor 10 TR Transformer 11 RE Relays 12 HE Heating element 13 F2 Fuses 50mA ...

Page 14: ...USER and TECHNICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 14 of 16 ...

Page 15: ...USER and TECHNICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 15 of 16 ...

Page 16: ...ICAL MANUAL KANMED Warming Cabinet GE 2 1370 Page 16 of 16 KanMed AB Gårdsfogdevägen 18B S 168 66 BROMMA SWEDEN Tel 46 0 8 564 80 630 Fax 46 0 8 564 80 639 E Mail info KANMED se Home page www KANMED se DISTRIBUTED BY ...

Reviews: