Technical manual
KANMED Operatherm OP3
Page
11
of
34
2. Fit the USB-memory stick containing the Service Software
(..or the user software) into the internal USB connector.
3. Connect the mains power, and within 5 seconds, press and
hold left AND right arrow buttons pressed simultaneously
4. When the message “Updating SW..” is displayed, the left-
and right arrow buttons can be released.
The “Ok” message indicate that the new software has been
successfully loaded.
5. When the Stand-by screen is displayed, disconnect the
mains power and remove the USB-memory stick from the
Control unit. Reconnect the mains power and let the system
boot as normal. The new software has now been loaded into
the system.
The software type is identified by the version mark in the
bottom right corner of the display.