If no PAD is connected, three log files will be generated (above). Four if a PAD is connected
(below).
The log files are in txt format and can be opened with a text editor, e.g. WordPad or similar. The
log files are mainly intended for factory level post analyze of any “un-explainable” errors in the CU
or the PAD.
Data example from the CU-session log file D-OPERAT:
Log: 0
Start time: Fri Jul 11 14:26:03 2008
End Time: Fri Jul 11 14:26:16 2008
Time: 13
Total Time: 13
Sessions: 1
Pad: , Mode: 0
Energy: 0, Max Temp: 26.06, Min Temp: 24.78
Alarm: 55, Fri Jul 11 14:26:16 2008
Log: 1
Start time: Fri Jul 11 14:27:56 2008
End Time: Fri Jul 11 14:28:52 2008
Time: 56
Total Time: 69
Sessions: 2
Pad: , Mode: 0
Energy: 2, Max Temp: 29.60, Min Temp: 26.94
Alarm: 60, Fri Jul 11 14:28:52 2008
Etc.
Data example from the PAD log file P-“PAD S/N”:
Log: 0
time: Tue Sep 8 14:34:40 2009