Control Unit Calibration Check
Note
: This operation requires the use of a Calibration Adapter
(part No. OP3-099).
Connect the calibration adapter (part No: OP3-099) to the
control unit and press on/off. The unit will automatically enter
Calibration Check mode.
The unit will display the measured resistance for each of the 8
channels. Channel 3 – 6 will display two sets of figures. The
reason for this is that channel 1 – 6 is measured with A/D -
converter 1, and channel 3 – 8 with A/D-converter 2. Channel
3-6 will therefore show the readings from both converters.
After the unit has returned steady and reliable values (after 2-
5 min), the readings (the measured resistance for each
channel) should correspond to the adapters reference
resistance (displayed on top of the screen) within 0,03 ohm. If
the error is larger than this, it will display a red marking over
the reading, otherwise a green.
Press MENU to exit.
NOTE
:
If one or several of the readings exceeds 0.03 ohm, leave the unit running for at least 10-15
minutes and check again. If one (or several) of the reading is still just outside limits, pls. restart
the control unit and perform the test once more. If the error is consistent, the control unit might
need a re-calibration. This function is also available in the Customer Software version.
Log Data Menu
Select Main MENU/Log Data
NOTE
:
This function is also available in the Customer Software version.
Control Unit Log
Select “Control Unit Log” and press “Enter” to display the CU
logs. Press [<] or [>] to step through the different logs. Press
“MENU” to exit.
NOTE
:
This function is also available in the Customer Software version.