6 Cleaning, Checks before use and Maintenance
General information
The WarmCloud control unit, hoses and remote control should be carefully cleaned between each
patient.
Follow the routines set forth by the person responsible for cleaning and disinfection in your hospital
and department and according to the instructions below.
Inspect all parts for wear and damages before and after cleaning.
Below is a list of cleaning and disinfection agents recommended by Kanmed. Using other agents
and/or exaggerated cleaning may reduce the lifetime of some or all parts. Contact Kanmed with
the product specification if in doubt.
Do not use bleach and solvents.
Kanmed recommends using the Kanmed Tube Protection Art no 700-0882
Cleaning:
Use soap water in a moist cloth.
Disinfection:
The WarmCloud, Remote control with cable and hose set withstands most common
surface disinfectants used in hospitals. Kanmed recommends DAX surface disinfectant which is
alcohol based. Do not use alcohol based solutions stronger than 70%
If in doubt, try on a small surface or contact your distributor or Kanmed.
Hose protection
Kanmed Disposable Tube Protection Art no 700-0882 is available to facilitate keeping the hoses
clean.
WarmCloud Mattresses
The Mattress is a single patient use disposable item and shall not be reused.
Checks before use
Check that all parts are clean and in good condition. Check that the hoses are properly connected
to the unit by pulling gently. Check that the unit performs a self-test at start. If Error Codes appear
and cannot be eliminated with the suggestions given in section 10 then you shall not use the unit.
Note: If the Main Unit or hose set has received mechanical damage, it should always be checked
and correct system function verified by technical staff.
Maintenance
Kanmed WarmCloud does not require any special maintenance apart from annual check according
to the service manual.
7 Accessories and Spare Parts
Article no
Product
Comment
9000
Kanmed WarmCloud
Complete system with standard accessories. 230V
9000-120
Kanmed WarmCloud
Complete system with standard accessories. 120V
OT-600-222
Warm-Air Mattress,
Wide
Single-Use Mattress with ties, Length 200 cm, Width 55 cm.
Delivered in box of 10 pcs
OT-600-211
Warm-Air Mattress,
Small
For “leg
-
up” positions and for Children
Single-Use Mattress with ties, Length 110 cm, Width 50 cm.
Delivered in box of 10 pcs
700-0882
Tube protection
Disposable thin plastic tube
9300
Remote control
OT-600-705 Remote control cable
9131
Wheel base
Wheel base with tube storage.
WC2-10
Hose set complete
Kit with two tubes
WC2-070
Manual
English Manual.
9079
Manual
Service manual English.
Test box
For service checks only
Other mattress sizes and accessories may be available Always check with your dealer or
www.kanmed.se
for the latest information.