User manual WC
Page
14 of 20
8 Technical information
Safety systems
Kanmed WarmCloud is equipped with several safety functions to prevent dangerous situations.
If a fault is detected an alarm will sound and an error code and symbol will appear on the display.
A code will appear on the Remote Control. Full description of Alarm functions and codes are found
in section 10.
Full description of the safety systems and functions are found in the service manual.
9 Technical Data
Voltage art no 9000
220 - 240 VAC 50/60 Hz
Voltage art no 9000-
120
100-130 VAC 50/60 Hz
Power consumption
230V:150 VA (Average under normal conditions), 700 VA (Peak value)
120V:150 VA (Average under normal conditions), 950 VA (Peak value)
Fuses
See service manual.
Inflation time
About 90s with OT-600-222 Mattress depending on inlet pressure.
Warming up time
About 10 minutes from room temperature (18-22°C) to 42°C.
Sound level alarm
65 dBA
Measurements
Main Unit: 310 x 295 x 370 mm.
Length of Air Hoses: 1500 mm.
Remote control cable length: 2900 mm.
Gas Connection
Medical air or apparatus air. 400.700kPa.
Weight Main Unit
15 kg incl. wheelbase and
tubing’s
.
Temperature
selection
31°C - 42°C, step of 1°C. Pre- 2°C.
Max contact
temperature
45°C on the mattress.
Pressure selection
0
–
50 mbar indicated as steps 1 to 9. One step is about 5 mbar.
Pre- 5mb.
Mechanical overpressure valve release at 75 mbar.
Patient applied part
Mattress.
Function supervision
Audio and visual alert. Error codes displayed on the control unit and
Remote Control.
Mattresses material
EVA plastic with non-woven cover.
Control unit and
mattresses.
Environmental data
Use
+15°C to +35°C during operation, less than 90% Relative Humidity
700 to 1060 hPa (equivalent to a max height of about 3000m.)
Control Unit
Environmental data
Storing
0°C to +40°C non-condensing.
Air pressure not relevant.
Control Unit
Environmental data
Transport
- 40°C to +40°C, non-condensing.
Air pressure not relevant.
Mattresses
Storing and transport
-40°C to +30°C.
Air pressure not relevant.
Modifications
Any modifications on the control unit, hand unit, hoses and the mattress
will void Kanmed
’s responsibilities totally
and are not allowed without
the written consent of Kanmed.
Creating Systems
Any person connecting the WarmCloud to any other systems mains
socket may have created a “system”
according to IEC 60601-1 section
16 and must be qualified to evaluate any consequences that may harm
staff, patient and equipment.
EU Classification
Control unit: MDD Class IIb.
All Mattresses: MDD Class I.
Expected Lifetime
Control unit WarmCloud.
Kanmed warrants a safe lifetime for the
Control unit of 10 years from first day of use. This is under the condition
that the unit has been maintained and serviced according to the user
and or service manual and that the unit has not been modified,
damaged or changed in any way or for any reason.