connector
lock
button
3720-06 Analog output terminal
The analog output pins are described below.
terminal
number
3720-06 Signal
name
Function
Descript
ion
1
ALM.ch1
Channel 1 measurement particle count or concentration
output upon alarm detection
terminal. The contact specification for this terminal is 60 V
0.4 A or less. This
Use within the range of
2
ALM.ch2
Channel 2 measurement particle count or
concentration output upon alarm detection
terminal. The contact specification for this terminal
is 60 V 0.4 A or less. Use within this range.
3
ALM.ch3
Channel 3 measurement particle count or
concentration output upon alarm detection
terminal. The contact specification for this terminal
is 60 V 0.4 A or less. Use within this range.
4
ALM.ch4
Channel 4 measurement particle count or
concentration output upon alarm detection
terminal. The contact specification for this terminal
is 60 V 0.4 A or less. Use within this range.
5
ALM.COM
Minus common terminal of alarm output resistor.
6
A-GND
Minus common terminal for current output.
7
Power(15-
28V)
DC power "+" connection terminal.
8
P-GND
DC power "GND" connection pin.