20
Troubleshooting
§6
Symptoms
Possible causes(s)
→
Solution(s)
Reference
The display does not
appear when the power is
turned ON.
The AC adapter is not connected properly.
→
Confirm the AC adapter and power cable.
Low battery
→
Replace the batteries.
→
Recharge the batteries.
6
The battery drains fast.
The battery is deteriorated.
→
Replace the battery
6
Particle count or
concentration is too high.
The actual concentration is high or the instrument may
malfunction.
→
Attach the provided Zero count filter and confirm
that the reading drops to zero. If the reading
remains high, the instrument may malfunction.
Please contact Kanomax USA.
7
Particle count or
concentration is too low.
Laser power failure or flow error
→
The instrument may malfunction.
Please contact Kanomax USA.
-
The printer does not
operate.
Incorrect baud rate setting
→
Confirm the printer setting.
-
Data can not be stored.
Saving data function is OFF
→
Confirm the save setting.
The number of recorded data may exceed 10,000.
12
10
The measured value of the
particles maybe higher in
Wi-Fi remote mode.
If Wi-Fi remote connection is used above a metal
conductor, the Wi-Fi radio wave may be reflected and affect
the measured result.
→
When using Wi-Fi connection above any metal
conductor, please use a non - metallic spacer of 3 cm or
more.
-