NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
16
Version 1.0.1
5.
Remove the protective nut and plug from the
UPW Inlet
fitting. Keep this
Flaretek nut and plug for use when moving or shipping the nebulizer.
(See shutdown instructions on page 87).
6.
Push the flared end of the tube onto the
UPW Inlet
.
7.
Slide the Flaretek nut into place and hand-tighten.
8.
Turn on the ultrapure water supply. Water then flows through the
instrument and out through the waste line. Ensure that the
UPW Inlet
has no leaks. If you see any leaks, tighten the fitting.
Connecting the Sample Inlet
To connect the Sample Inlet, follow these instructions:
1.
On the front panel, turn the Sample Selector Valve to DIRECT.
2.
Remove the 1/4-28 plug protective cap from the
Sample Inlet
fitting.
3.
Install the supplied 1/4-28 tube adapter to the
Sample Inlet
and hand
tighten.
4.
Connect the supplied peristaltic pump Tygon tubing to the adapter and
verify that the adapter is fully inserted into the pump tubing.
Flaretek Nut
Sample Selector
Valve
Sample Inlet
cap