NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
32
Version 1.0.1
3.
Snap the provided fill bottle into place in the bracket as shown in the
photo below.
4.
Insert the fitting on the end of the bottle’s lower tube into the
n-Butanol
Supply
inlet.
5.
Insert the fitting on the end of the bottle’s upper tube into the
n-Butanol
Exhaust
outlet.
6.
Carefully pour n-Butanol into the fill bottle. Do not fill past the n-Butanol
exhaust fitting level. The n-Butanol will not flow into the CPC until the
connections are made, the instrument is powered on, and the warm-up
cycle is complete.
Warning:
n-Butanol (n-butyl alcohol) is flammable and toxic; follow
appropriate chemical handling procedures.
Connecting the Exhaust
The air flow containing n-Butanol vapor exits the Fast CPC from the Exhaust
port. When operating the Fast CPC in a confined space, vent the exhaust
n-Butanol
Exhaust
n-Butanol
Supply