NanoParticle Nebulizer: Operation
Liquid NanoParticle Sizer User Guide
43
Version 1.0.1
5.
Use one supplied Ethernet cable to connect the Ethernet switch to a
nearby network access point, or to the computer with Kanolysis Model
S330 installed. Alt
hough it is possible to use the computer’s built
-in
Ethernet port if equipped, Kanomax FMT recommends using a dedicated
USB-Ethernet adapter, supplied in the LNS Accessory Kit, to avoid
interfering with connectivity to other networks.
6.
For IPv4 configurations, apply the network settings shown in Table 5. To
access the network settings screen of each instrument, press
F2
on the
display screen and touch
More Settings / Configuration
and
Network
buttons.
Table 5: IPv4 Configuration Settings
Device name
IP address
Network mask
Default
gateway
DNS
servers
Computer
10.41.4.1
255.255.255.0
(blank)
(blank)
NPN
10.41.4.2
255.255.255.0
0.0.0.0
N/A
AFIMC
10.41.4.3
255.255.255.0
0.0.0.0
N/A
FCPC
10.41.4.4
255.255.255.0
0.0.0.0
N/A
Note that if the computer is attached to other networks, the user must
also ensure there is no overlap with this subnet.